A Novel Methodology for the Evaluation of Adapt-to-Position Segment Shape Optimisation in MR-Guided Radiotherapy

Daily MR guided adaptive radiotherapy represents the current pinnacle of image guidance sophistication in external beam radiotherapy. In this work, we present a novel methodology for the quantitative assessment of the Elekta Unity MR Linac (Elekta AB, Stockholm, Sweden) Adapt to Position (ATP) Segment Shape Optimisation (SSO) planning algorithm as used in a clinical environment. MR-safe positioning devices manufactured via 3D printing were used to precisely offset a phantom in known increments along each axis of the IEC 61217 coordinate system. Ion chamber and radiochromic film (RCF) measurements were used to compare adaptive algorithm dosimetry and field repositioning at these offset positions compared to a phantom reference position. The registration read back 12 known spacer offsets within 0.6 mm. Without adaptation, the chamber detected offsets of 2 mm along x (3 mm on an older plan) and 3 mm along y and z. For 42 SSO adaptations (5–30 mm) the chamber read + 0.2% (SD 0.6%) against the calculated dose to water; 3 points lay outside ± 1.1% and one at 2.03 standard uncertainties of its own budget (k = 2). The plan-side gamma pass rate fell as the apertures departed from a pure translation (rho = − 0.81, 33 plans); the chamber did not resolve this. All film comparisons passed a global 2%/2 mm gamma criterion. It may be concluded that the ATP SSO algorithm delivered the adapted calculated dose within the per-point uncertainty at 41 of 42 points, and that the chamber alone cannot resolve errors below 2 to 3 mm.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-23
DOI
https://doi.org/10.21203/rs.3.rs-11688388/v1
Primary Topic
Advanced Radiotherapy Techniques
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

A Novel Methodology for the Evaluation of Adapt-to-Position Segment Shape Optimisation in MR-Guided Radiotherapy

Rick D. Franich, Marchant Van der Walt, John A. Baines
Research Square
Advanced Radiotherapy Techniques
preprint

A Novel Methodology for the Evaluation of Adapt-to-Position Segment Shape Optimisation in MR-Guided Radiotherapy

Rick D. Franich, Marchant Van der Walt, John A. Baines
preprint en

Abstract

Daily MR guided adaptive radiotherapy represents the current pinnacle of image guidance sophistication in external beam radiotherapy. In this work, we present a novel methodology for the quantitative assessment of the Elekta Unity MR Linac (Elekta AB, Stockholm, Sweden) Adapt to Position (ATP) Segment Shape Optimisation (SSO) planning algorithm as used in a clinical environment. MR-safe positioning devices manufactured via 3D printing were used to precisely offset a phantom in known increments along each axis of the IEC 61217 coordinate system. Ion chamber and radiochromic film (RCF) measurements were used to compare adaptive algorithm dosimetry and field repositioning at these offset positions compared to a phantom reference position. The registration read back 12 known spacer offsets within 0.6 mm. Without adaptation, the chamber detected offsets of 2 mm along x (3 mm on an older plan) and 3 mm along y and z. For 42 SSO adaptations (5–30 mm) the chamber read + 0.2% (SD 0.6%) against the calculated dose to water; 3 points lay outside ± 1.1% and one at 2.03 standard uncertainties of its own budget (k = 2). The plan-side gamma pass rate fell as the apertures departed from a pure translation (rho = − 0.81, 33 plans); the chamber did not resolve this. All film comparisons passed a global 2%/2 mm gamma criterion. It may be concluded that the ATP SSO algorithm delivered the adapted calculated dose within the per-point uncertainty at 41 of 42 points, and that the chamber alone cannot resolve errors below 2 to 3 mm.

Research Square
Townsville Hospital (AU), RMIT University (AU)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon