Fit-for-purpose bioanalytical validation of a three-cytokine human PBMC ELISpot panel (IFN-γ, IL-2, IL-4) using CEFX and CPI antigen pools for clinical trial immunomonitoring under a GLP-aligned quality system

Cellular immunity is an increasingly required endpoint in clinical trials of vaccines, immuno-oncology agents, and cell therapies. Multi-cytokine validation packages under a single quality system remain uncommon, and the permissible interval between blood draw and plating or freezing is rarely demonstrated experimentally. We performed GLP-aligned, fit-for-purpose validation of three human peripheral blood mononuclear cell (PBMC) ELISpot assays, following Global CRO Council recommendations and FDA, NMPA, and CFDI guidance. IFN-gamma and IL-2 were stimulated with CEFX Ultra SuperStim Pool (24 h each, plus a validated 48 h IFN-gamma regimen) and IL-4 with CPI Antigen Pool (CTL-CPI-001; 48 h) throughout all validation arms. We characterized limit of detection (LOD), lower limit of quantification (LLOQ), precision, linearity across six plating densities, robustness, pre-analytical PBMC functional stability under five delayed-processing conditions in fresh and cryopreserved arms. All assays met predefined criteria (>=80% of samples with %CV <=40% above LLOQ; R^2 >=0.97); LOD/LLOQ were 13/20 and 37/45 SFU/well for IFN-gamma at 24 h and 48 h, 9/27 for IL-2, and 4/27 for IL-4. Freshly plated PBMC remained within acceptance after up to 8 h plus 24 h of ambient holding, whereas cryopreservation after prolonged delay failed at 8+16 h for IL-2 and IL-4 and at 8+24 h for all three assays, driven by post-thaw viability less than 80%. These data define evidence-based sample-handling windows for multicenter trials. The resulting three-cytokine platform is fit-for-purpose for regulated clinical immunomonitoring.

Authors

Publication Details

Journal
medRxiv
Published
2026-11-23
DOI
https://doi.org/10.64898/2026.11.22.26368554
Primary Topic
Biosimilars and Bioanalytical Methods
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Fit-for-purpose bioanalytical validation of a three-cytokine human PBMC ELISpot panel (IFN-γ, IL-2, IL-4) using CEFX and CPI antigen pools for clinical trial immunomonitoring under a GLP-aligned quality system

Min Tang, Zhuo Wang, Peng Guo, Ning Zhang et al.
medRxiv
Biosimilars and Bioanalytical Methods
preprint

Fit-for-purpose bioanalytical validation of a three-cytokine human PBMC ELISpot panel (IFN-γ, IL-2, IL-4) using CEFX and CPI antigen pools for clinical trial immunomonitoring under a GLP-aligned quality system

Min Tang, Zhuo Wang, Peng Guo, Ning Zhang, Pengyang Fan, Lijie Jiang, Xiaoliang Zhai, Kaitong Li, Zhenyuan Cui, Hongjuan Wang, Hongxu Cao
preprint en

Abstract

Cellular immunity is an increasingly required endpoint in clinical trials of vaccines, immuno-oncology agents, and cell therapies. Multi-cytokine validation packages under a single quality system remain uncommon, and the permissible interval between blood draw and plating or freezing is rarely demonstrated experimentally. We performed GLP-aligned, fit-for-purpose validation of three human peripheral blood mononuclear cell (PBMC) ELISpot assays, following Global CRO Council recommendations and FDA, NMPA, and CFDI guidance. IFN-gamma and IL-2 were stimulated with CEFX Ultra SuperStim Pool (24 h each, plus a validated 48 h IFN-gamma regimen) and IL-4 with CPI Antigen Pool (CTL-CPI-001; 48 h) throughout all validation arms. We characterized limit of detection (LOD), lower limit of quantification (LLOQ), precision, linearity across six plating densities, robustness, pre-analytical PBMC functional stability under five delayed-processing conditions in fresh and cryopreserved arms. All assays met predefined criteria (>=80% of samples with %CV <=40% above LLOQ; R^2 >=0.97); LOD/LLOQ were 13/20 and 37/45 SFU/well for IFN-gamma at 24 h and 48 h, 9/27 for IL-2, and 4/27 for IL-4. Freshly plated PBMC remained within acceptance after up to 8 h plus 24 h of ambient holding, whereas cryopreservation after prolonged delay failed at 8+16 h for IL-2 and IL-4 and at 8+24 h for all three assays, driven by post-thaw viability less than 80%. These data define evidence-based sample-handling windows for multicenter trials. The resulting three-cytokine platform is fit-for-purpose for regulated clinical immunomonitoring.

medRxiv
Biosimilars and Bioanalytical Methods
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Fit-for-purpose bioanalytical validation of a three-cytokine human PBMC ELISpot panel (IFN-γ, IL-2, IL-4) using CEFX and CPI antigen pools for clinical trial immunomonitoring under a GLP-aligned quality system — Min Tang, Zhuo Wang, et al. · medRxiv (2026) | TGRS Research Map | TGRS