Pumpless extracorporeal lung assist device in the management of idiopathic pulmonary arterial hypertension – implantation, weaning and explantation in a bridge-to-decision concept
Via a primary sternotomy approach, cardiopulmonary bypass is established in a standard fashion including a separate vent in the main pulmonary artery. The outlet cannula, an angled Berlin Heart EXCOR cannula, is implanted into the body of the small, underfilled, left atrium using felt-supported single sutures. The inlet cannula, a 12 mm Berlin Heart graft adapter cannula connected to a 12 mm Terumo prosthesis and draining deoxygenated blood from the main pulmonary artery, is tunnelled through the epigastric skin and subsequently anastomosed with the main pulmonary artery. A low-resistance oxygenator (QUADROX, Maquet) is used for paracorporeal lung assist.
Authors
- Jürgen Hörer (ORCID: https://orcid.org/0000-0002-2716-1424)
- Fabian Alexander Kari (ORCID: https://orcid.org/0000-0002-0985-9650)
- Nikolaus Haas
- Joseph Pattathu
- Sebastian G. Michel (ORCID: https://orcid.org/0000-0002-0841-860X)
- Christian P. Schneider (ORCID: https://orcid.org/0000-0003-4501-3094)
- A Maskos
Institutions
- LMU Klinikum (DE)
- German Centre for Cardiovascular Research (DE)
- German Center for Lung Research (DE)
- Deutsches Herzzentrum München (DE)
- TUM Klinikum (DE)
- Technical University of Munich (DE)
- Ludwig-Maximilians-Universität München (DE)
Publication Details
- Journal
- Multimedia Manual of Cardio-Thoracic Surgery
- Published
- 2026-11-23
- DOI
- https://doi.org/10.1510/mmcts.2026.079
- Primary Topic
- Mechanical Circulatory Support Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.00