Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium–zinc–telluride and NaI(Tl) detectors

BACKGROUND: Lutetium-177 (177Lu) is a promising theranostic radionuclide. Accurate quantitative SPECT imaging is essential for 177Lu patient-specific dosimetry. Although the 113 keV photopeak has a higher scatter fraction, it matches the high energy resolution of cadmium-zinc-telluride (CZT) detectors and avoids 208 keV limitations. However, whether 113 keV-only imaging provides sufficient lesion detectability remains unclear. Focusing on the 113 keV photopeak, this study compared two dual-head SPECT/computed tomography systems: conventional NaI(Tl) scintillators with low-energy high-resolution collimators, and CZT semiconductor detectors with wide-energy high-resolution collimators. This study compares these technologies and establishes optimized reconstruction protocols. METHODS: National Electrical Manufacturers Association/International Electrotechnical Commission phantoms were scanned on both systems using the 113 keV photopeak. Ordered subset expectation maximization reconstructions were performed (2-12 iterations) with Gaussian [full-width-at-half-maximum (FWHM): 1.5-4.0 mm) or Butterworth filters (cut-off frequency: 0.25-0.50 cycle/pixel). Attenuation, scatter, and resolution corrections were applied. Image quality was quantified using background coefficient of variation (COV_B), contrast recovery coefficient (CRC), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR). RESULTS: CZT delivered superior image quality, achieving higher CRC, CNR, and SNR with lower COV_B than NaI(Tl). Optimal parameters were 2 iterations, 10 subsets, Gaussian FWHM 3.0 mm, and Butterworth cut-off 0.35 cycles/pixel (CZT) and 0.30 cycles/pixel [NaI(Tl)]. These settings yielded robust performance across target-to-background ratios of 8 : 1, 16 : 1, and 32 : 1. CONCLUSION: For the 177Lu 113 keV peak, CZT detectors markedly enhance SPECT imaging, improving lesion detectability and quantification. Tailored reconstruction optimization further strengthens image quality and reliability, supporting precision dosimetry and CZT adoption in radionuclide therapy.

Authors

Institutions

Publication Details

Journal
Nuclear Medicine Communications
Published
2026-11-23
DOI
https://doi.org/10.1097/mnm.0000000000002275
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium–zinc–telluride and NaI(Tl) detectors

Qiang Jia, Hongjin Zhou, Wang Shen, Wei Zheng et al.
Nuclear Medicine Communications
Medical Imaging Techniques and Applications
article

Optimisation of reconstruction parameters for lutetium-177 SPECT imaging: comparative performance of cadmium–zinc–telluride and NaI(Tl) detectors

Qiang Jia, Hongjin Zhou, Wang Shen, Wei Zheng, Ruyi Zhang
article en

Abstract

BACKGROUND: Lutetium-177 (177Lu) is a promising theranostic radionuclide. Accurate quantitative SPECT imaging is essential for 177Lu patient-specific dosimetry. Although the 113 keV photopeak has a higher scatter fraction, it matches the high energy resolution of cadmium-zinc-telluride (CZT) detectors and avoids 208 keV limitations. However, whether 113 keV-only imaging provides sufficient lesion detectability remains unclear. Focusing on the 113 keV photopeak, this study compared two dual-head SPECT/computed tomography systems: conventional NaI(Tl) scintillators with low-energy high-resolution collimators, and CZT semiconductor detectors with wide-energy high-resolution collimators. This study compares these technologies and establishes optimized reconstruction protocols. METHODS: National Electrical Manufacturers Association/International Electrotechnical Commission phantoms were scanned on both systems using the 113 keV photopeak. Ordered subset expectation maximization reconstructions were performed (2-12 iterations) with Gaussian [full-width-at-half-maximum (FWHM): 1.5-4.0 mm) or Butterworth filters (cut-off frequency: 0.25-0.50 cycle/pixel). Attenuation, scatter, and resolution corrections were applied. Image quality was quantified using background coefficient of variation (COV_B), contrast recovery coefficient (CRC), contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR). RESULTS: CZT delivered superior image quality, achieving higher CRC, CNR, and SNR with lower COV_B than NaI(Tl). Optimal parameters were 2 iterations, 10 subsets, Gaussian FWHM 3.0 mm, and Butterworth cut-off 0.35 cycles/pixel (CZT) and 0.30 cycles/pixel [NaI(Tl)]. These settings yielded robust performance across target-to-background ratios of 8 : 1, 16 : 1, and 32 : 1. CONCLUSION: For the 177Lu 113 keV peak, CZT detectors markedly enhance SPECT imaging, improving lesion detectability and quantification. Tailored reconstruction optimization further strengthens image quality and reliability, supporting precision dosimetry and CZT adoption in radionuclide therapy.

Nuclear Medicine Communications
Tianjin Medical University General Hospital (CN)
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon