A portable evaporative preconcentration platform integrated with continuous real-time UV spectroscopy

In many practical scenarios, analyte concentrations in aqueous samples are too low for direct detection, creating a need for portable on-site preconcentration. We developed a portable evaporative preconcentration platform that uses an integrated air pump to remove water while retaining nonvolatile or low-volatility solutes and combines real-time temperature and relative humidity sensing with continuous flow-through ultraviolet (UV) spectroscopy. The evaporation ratio varied linearly with the relative humidity increase of the sweeping gas, which flows through the preconcentrator to remove evaporated solvent. This increase was calculated as the relative humidity after the preconcentrator minus that before it, with R 2 values of 0.98 and 0.96 at liquid flow rates of 5 and 10 μL/min, respectively. Poly(sodium 4-styrenesulfonate) was used as a model analyte, whose concentration was quantified using the absorbance peak height at 262 nm, which responded linearly from 0.1 to 1.0 mg/mL ( R 2 = 0.998). Continuous experiments demonstrated predictable steady-state transitions following changes in liquid flow rate and operating temperature. The downstream fluidic volume was measured directly as approximately 150 μL and independently estimated from the transition-time data as 146 ± 32 μL, indicating that the response time was primarily volume limited. Under ambient and laboratory-controlled conditions, measured concentration factors ranged from 1.16 to 2.58. Estimates incorporating time-resolved sensor measurements and operating parameters provided closer agreement with the measured values in several tests than predictions based on fixed operating conditions, although discrepancies remained in some tests. These results demonstrate a compact platform for continuous analyte enrichment and quantitative monitoring under variable operating conditions.

Authors

Institutions

Publication Details

Journal
Analytical and Bioanalytical Chemistry
Published
2026-11-23
DOI
https://doi.org/10.1007/s00216-026-06922-3
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A portable evaporative preconcentration platform integrated with continuous real-time UV spectroscopy

Tanya Hutter, Yufeng Su, Vanessa Do
Analytical and Bioanalytical Chemistry
Analytical chemistry methods development
article

A portable evaporative preconcentration platform integrated with continuous real-time UV spectroscopy

Tanya Hutter, Yufeng Su, Vanessa Do
article en

Abstract

In many practical scenarios, analyte concentrations in aqueous samples are too low for direct detection, creating a need for portable on-site preconcentration. We developed a portable evaporative preconcentration platform that uses an integrated air pump to remove water while retaining nonvolatile or low-volatility solutes and combines real-time temperature and relative humidity sensing with continuous flow-through ultraviolet (UV) spectroscopy. The evaporation ratio varied linearly with the relative humidity increase of the sweeping gas, which flows through the preconcentrator to remove evaporated solvent. This increase was calculated as the relative humidity after the preconcentrator minus that before it, with R 2 values of 0.98 and 0.96 at liquid flow rates of 5 and 10 μL/min, respectively. Poly(sodium 4-styrenesulfonate) was used as a model analyte, whose concentration was quantified using the absorbance peak height at 262 nm, which responded linearly from 0.1 to 1.0 mg/mL ( R 2 = 0.998). Continuous experiments demonstrated predictable steady-state transitions following changes in liquid flow rate and operating temperature. The downstream fluidic volume was measured directly as approximately 150 μL and independently estimated from the transition-time data as 146 ± 32 μL, indicating that the response time was primarily volume limited. Under ambient and laboratory-controlled conditions, measured concentration factors ranged from 1.16 to 2.58. Estimates incorporating time-resolved sensor measurements and operating parameters provided closer agreement with the measured values in several tests than predictions based on fixed operating conditions, although discrepancies remained in some tests. These results demonstrate a compact platform for continuous analyte enrichment and quantitative monitoring under variable operating conditions.

Analytical and Bioanalytical Chemistry
Texas Materials Institute (US), The University of Texas at Austin (US)
University of Texas at Austin
Analytical chemistry methods development
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon