Same Model, Different Patient: Per Patient Decision Changes When a Chest Radiograph Model Is Redeployed
Chest radiograph models are often redeployed (new DICOM decoder, resize library, precision or inference engine) after a check that an aggregate metric has not changed. We reran four public models (a PSPNet cardiothoracic ratio (CTR), TorchXRayVision DenseNet121 at its library default thresholds, and RAD-DINO and BiomedCLIP opacity classifiers) under such changes on MIDRC-RICORD-1C (360 patients, one frontal image each), replicated on COVID-19-NY-SBU (700 patients), and compared each patient flip rate with 20 benign input perturbations. Replacing an antialiased resize passed an equivalence test on the mean CTR at a post hoc margin of 0.02, yet the ratio moved by more than 0.02 in 26.1% (95% CI 21.6 to 31.0) of patients (22.2% after excluding implausible masks, 16.9% on PA views), against 7.0% under a one pixel shift. A mean test speaks only about the average patient, so the quantity that matches a decision is the share of patients who flip, read against such a baseline. For DenseNet121 after Holm correction, 6 of 15 distinct effects flip more patients than the strongest perturbation (4.2%) on RICORD and 4 on SBU: resizes without antialiasing, the VOI LUT, two misconfigurations and the worst of five INT8 builds, a build picked after the data were seen. At thresholds refit on half the patients the resizes no longer reach significance. Five INT8 builds per calibration source flipped 2.2% to 8.9% and 1.7% to 3.4% of 179 held out patients, and after Holm 0 SBU and 0 RICORD calibrated builds exceeded the perturbations (4 and 4 on a T4). Seven fp32 engine changes flip no patient. SBU has no labels, MONOCHROME1 or VOI LUT files, so it replicates the CTR and resize results only; all RICORD analyses are exploratory and all runs but a T4 rerun use one GPU.
Authors
- Sharven Rane (ORCID: https://orcid.org/0009-0008-4800-690X)
Publication Details
- Journal
- Zenodo (CERN European Organization for Nuclear Research)
- Published
- 2026-11-22
- DOI
- https://doi.org/10.5281/zenodo.24433996
- Primary Topic
- COVID-19 diagnosis using AI
- Type
- preprint