Same Model, Different Patient: Per Patient Decision Changes When a Chest Radiograph Model Is Redeployed

Chest radiograph models are often redeployed (new DICOM decoder, resize library, precision or inference engine) after a check that an aggregate metric has not changed. We reran four public models (a PSPNet cardiothoracic ratio (CTR), TorchXRayVision DenseNet121 at its library default thresholds, and RAD-DINO and BiomedCLIP opacity classifiers) under such changes on MIDRC-RICORD-1C (360 patients, one frontal image each), replicated on COVID-19-NY-SBU (700 patients), and compared each patient flip rate with 20 benign input perturbations. Replacing an antialiased resize passed an equivalence test on the mean CTR at a post hoc margin of 0.02, yet the ratio moved by more than 0.02 in 26.1% (95% CI 21.6 to 31.0) of patients (22.2% after excluding implausible masks, 16.9% on PA views), against 7.0% under a one pixel shift. A mean test speaks only about the average patient, so the quantity that matches a decision is the share of patients who flip, read against such a baseline. For DenseNet121 after Holm correction, 6 of 15 distinct effects flip more patients than the strongest perturbation (4.2%) on RICORD and 4 on SBU: resizes without antialiasing, the VOI LUT, two misconfigurations and the worst of five INT8 builds, a build picked after the data were seen. At thresholds refit on half the patients the resizes no longer reach significance. Five INT8 builds per calibration source flipped 2.2% to 8.9% and 1.7% to 3.4% of 179 held out patients, and after Holm 0 SBU and 0 RICORD calibrated builds exceeded the perturbations (4 and 4 on a T4). Seven fp32 engine changes flip no patient. SBU has no labels, MONOCHROME1 or VOI LUT files, so it replicates the CTR and resize results only; all RICORD analyses are exploratory and all runs but a T4 rerun use one GPU.

Authors

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2026-11-22
DOI
https://doi.org/10.5281/zenodo.24433996
Primary Topic
COVID-19 diagnosis using AI
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Same Model, Different Patient: Per Patient Decision Changes When a Chest Radiograph Model Is Redeployed

Sharven Rane
Zenodo (CERN European Organization for Nuclear Research)
COVID-19 diagnosis using AI
preprint

Same Model, Different Patient: Per Patient Decision Changes When a Chest Radiograph Model Is Redeployed

Sharven Rane
preprint en

Abstract

Chest radiograph models are often redeployed (new DICOM decoder, resize library, precision or inference engine) after a check that an aggregate metric has not changed. We reran four public models (a PSPNet cardiothoracic ratio (CTR), TorchXRayVision DenseNet121 at its library default thresholds, and RAD-DINO and BiomedCLIP opacity classifiers) under such changes on MIDRC-RICORD-1C (360 patients, one frontal image each), replicated on COVID-19-NY-SBU (700 patients), and compared each patient flip rate with 20 benign input perturbations. Replacing an antialiased resize passed an equivalence test on the mean CTR at a post hoc margin of 0.02, yet the ratio moved by more than 0.02 in 26.1% (95% CI 21.6 to 31.0) of patients (22.2% after excluding implausible masks, 16.9% on PA views), against 7.0% under a one pixel shift. A mean test speaks only about the average patient, so the quantity that matches a decision is the share of patients who flip, read against such a baseline. For DenseNet121 after Holm correction, 6 of 15 distinct effects flip more patients than the strongest perturbation (4.2%) on RICORD and 4 on SBU: resizes without antialiasing, the VOI LUT, two misconfigurations and the worst of five INT8 builds, a build picked after the data were seen. At thresholds refit on half the patients the resizes no longer reach significance. Five INT8 builds per calibration source flipped 2.2% to 8.9% and 1.7% to 3.4% of 179 held out patients, and after Holm 0 SBU and 0 RICORD calibrated builds exceeded the perturbations (4 and 4 on a T4). Seven fp32 engine changes flip no patient. SBU has no labels, MONOCHROME1 or VOI LUT files, so it replicates the CTR and resize results only; all RICORD analyses are exploratory and all runs but a T4 rerun use one GPU.

Zenodo (CERN European Organization for Nuclear Research)
COVID-19 diagnosis using AI
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Same Model, Different Patient: Per Patient Decision Changes When a Chest Radiograph Model Is Redeployed — Sharven Rane · Zenodo (CERN European Organization for Nuclear Research) (2026) | TGRS Research Map | TGRS