A testable surface-strain prediction for a polymeric aortic valve leaflet under matched pressure
This dataset supports a computational study of how leaflet thickness changes surface strain in a nominal 23-mm polymeric aortic valve design. At a matched pressure of 100 Pa and with the nominal attachment geometry, reducing the thickness of the primary modeled cusp from 0.450 to 0.379 mm increased the regional, area-weighted positive maximum principal Green–Lagrange strain on its specified face by 454.49 microstrain. Increasing the attachment radius on the thin cusp alone by 0.50% reversed the sign of the predicted difference. Direct measurements of the predicted strain difference on fabricated candidate leaflets are pending.