Quantitative flow ratio (QFR) in routine practice: diagnostic performance and reproducibility

Background Quantitative flow ratio is an angiography-derived, adenosine- and wire-free method for functional coronary assessment. Data on its diagnostic performance and reproducibility relative to invasive indices remain limited. Methods We prospectively enrolled 95 patients with 101 coronary vessels. Instantaneous wave-free ratio was assessed in 76 vessels and resting full-cycle ratio in 25. Haemodynamic significance was defined as an invasive index of 0.89 or less and a quantitative flow ratio of 0.80 or less. Observer 1 performed two measurements and observer 2 one independent measurement. Diagnostic performance and reproducibility were assessed using intraclass correlation coefficients, root-mean-square within-subject coefficients of variation, Bland-Altman analysis, Cohen's kappa and classification changes. Results Vessels comprised 52 left anterior descending, 19 left circumflex and 30 right coronary arteries. Correlation coefficients were 0.722, 0.649 and 0.703, respectively (all p < 0.001). Corresponding areas under the receiver-operating-characteristic curve were 0.919, 0.930 and 0.899, with diagnostic accuracies of 89.1%, 88.1% and 88.1%. Intraobserver and interobserver intraclass correlation coefficients were 0.869 and 0.855. Mean differences were 0.018 ± 0.070 and 0.016 ± 0.075, with 95% limits of agreement from −0.120 to 0.156 and from −0.131 to 0.164. Three vessels were reclassified in each comparison. Agreement did not differ significantly between vessels with invasive indices of 0.89 or less and those with values above 0.89. Conclusion Quantitative flow ratio demonstrated good diagnostic performance and reproducibility. However, individual measurement differences may be clinically relevant near the 0.80 threshold. Borderline results warrant cautious interpretation and consideration of invasive physiological confirmation when treatment decisions may be affected.

Authors

Institutions

Publication Details

Journal
IJC Heart & Vasculature
Published
2026-11-21
DOI
https://doi.org/10.1016/j.ijcha.2026.102045
Primary Topic
Coronary Interventions and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Quantitative flow ratio (QFR) in routine practice: diagnostic performance and reproducibility

Joscha Kandels, Karsten Lenk, Marios Antoniadis, U. Laufs et al.
IJC Heart & Vasculature
Coronary Interventions and Diagnostics
article

Quantitative flow ratio (QFR) in routine practice: diagnostic performance and reproducibility

Joscha Kandels, Karsten Lenk, Marios Antoniadis, U. Laufs, J. Stader, M. Metze, L. Weissenseel, M. Blum, R. Wachter
article en

Abstract

Background Quantitative flow ratio is an angiography-derived, adenosine- and wire-free method for functional coronary assessment. Data on its diagnostic performance and reproducibility relative to invasive indices remain limited. Methods We prospectively enrolled 95 patients with 101 coronary vessels. Instantaneous wave-free ratio was assessed in 76 vessels and resting full-cycle ratio in 25. Haemodynamic significance was defined as an invasive index of 0.89 or less and a quantitative flow ratio of 0.80 or less. Observer 1 performed two measurements and observer 2 one independent measurement. Diagnostic performance and reproducibility were assessed using intraclass correlation coefficients, root-mean-square within-subject coefficients of variation, Bland-Altman analysis, Cohen's kappa and classification changes. Results Vessels comprised 52 left anterior descending, 19 left circumflex and 30 right coronary arteries. Correlation coefficients were 0.722, 0.649 and 0.703, respectively (all p < 0.001). Corresponding areas under the receiver-operating-characteristic curve were 0.919, 0.930 and 0.899, with diagnostic accuracies of 89.1%, 88.1% and 88.1%. Intraobserver and interobserver intraclass correlation coefficients were 0.869 and 0.855. Mean differences were 0.018 ± 0.070 and 0.016 ± 0.075, with 95% limits of agreement from −0.120 to 0.156 and from −0.131 to 0.164. Three vessels were reclassified in each comparison. Agreement did not differ significantly between vessels with invasive indices of 0.89 or less and those with values above 0.89. Conclusion Quantitative flow ratio demonstrated good diagnostic performance and reproducibility. However, individual measurement differences may be clinically relevant near the 0.80 threshold. Borderline results warrant cautious interpretation and consideration of invasive physiological confirmation when treatment decisions may be affected.

IJC Heart & VasculatureVol. 67
University Hospital Leipzig (DE), BG Klinikum Bergmannstrost Halle (DE)
Coronary Interventions and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon