Case study on thermal design and analysis of a sub-Kelvin cryogenic system for the CSNS-II ultracold neutron source
To support the Phase II ultracold neutron source at China Spallation Neutron Source, a sub-Kelvin cryogenic system capable of maintaining a 70 mm × 70 mm × 1000 mm superfluid helium converter below 0.7 K was designed and thermally analyzed. The system integrates a 1.4 K 4 He loop, a 0.5 K 3 He loop, and an ultrapure 4 He loop. Thermal loads from structural conduction, radiation, residual gas conduction, and process cooling were evaluated. The optimal 3 He inlet pressure was determined as 95 000 Pa to achieve balanced thermal load distribution. A spiral heat exchanger with a required tube length of 191 mm was designed for 2.6 K precooling. The minimum required 3 He– 4 He interfacial heat transfer area was calculated as 456 cm 2 . A porous alumina superleak achieved a 3 He/ 4 He isotopic ratio of 10 −10 . Four GM cryocoolers satisfy the total cooling demand of 89.49 W at 45 K and 1.958 W at 3.05 K.
Authors
- Meiying Ding
- Yu Yang (ORCID: https://orcid.org/0000-0002-2159-7382)
- Tiancheng Yi (ORCID: https://orcid.org/0000-0003-1955-0224)
- Haodong Wang (ORCID: https://orcid.org/0000-0002-4460-7758)
- Kun He (ORCID: https://orcid.org/0000-0001-8943-8671)
- Yijie Cai
- Jianzhen Li
- Yang Meng
- Songlin Wang
- Dongdong Wang
- Yu Zhang
- Dongdong Zhou
- Tianjiao Liang
- Qidong He
- Xuefen Han
Institutions
- Chinese Academy of Sciences (CN)
- China Spallation Neutron Source (CN)
- Institute of High Energy Physics (CN)
- Xi'an Jiaotong University (CN)
Publication Details
- Journal
- Case Studies in Thermal Engineering
- Published
- 2026-11-21
- DOI
- https://doi.org/10.1016/j.csite.2026.108741
- Primary Topic
- Spacecraft and Cryogenic Technologies
- Type
- article
- Field-Weighted Citation Impact
- 0.00