Mismatch Between Rotational and Linear Speed Affects Primary Stability of Dental Implants: An Experimental Study

Introduction Rotational and linear insertion speeds must be synchronized to ensure proper dental implant advancement. This study investigated the effects of a mismatch between rotational and linear speeds during implant insertion on primary stability. Methods Thirty identical dental implants were inserted into synthetic and natural bone at a constant rotational speed of 25 rpm. Correct insertion (CI) was defined as synchronized rotational and linear speeds. Accelerated insertion (AI) referred to a higher linear speed relative to the rotational speed. Primary stability was evaluated using maximum insertion torque (MIT), implant stability quotient (ISQ), and push-in test parameters. Results In natural bone, correct insertion produced significantly higher primary stability than accelerated insertion, as measured by MIT ( P < .001), ISQ ( P = .013), bone–implant construct stiffness ( P = .019), and maximum push-in force ( P < .001). In synthetic bone, no significant differences in stability indicators were found between insertion protocols, but axial force in the AI group was significantly higher compared to the CI group ( P = .013). For correct insertion in natural bone, a positive correlation was found between MIT and maximum push-in force (ρ = 0.867, P = .015). In contrast, no other correlation among stability indicators was statistically significant. Conclusion Mismatch between linear and rotational insertion speeds was associated with reduced primary stability in natural bone and increased axial force in synthetic bone. Clinical relevance Monitoring axial force during implant insertion, particularly in low-density bone, may provide valuable feedback on insertion accuracy, but requires further investigation.

Authors

Institutions

Publication Details

Journal
International Dental Journal
Published
2026-11-20
DOI
https://doi.org/10.1016/j.identj.2026.111345
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Mismatch Between Rotational and Linear Speed Affects Primary Stability of Dental Implants: An Experimental Study

Gholamreza Rouhi, Mahdi Farrokhi Kashtiban
2 links
International Dental Journal
Dental Implant Techniques and Outcomes
article

Mismatch Between Rotational and Linear Speed Affects Primary Stability of Dental Implants: An Experimental Study

Gholamreza Rouhi, Mahdi Farrokhi Kashtiban
article en

Abstract

Introduction Rotational and linear insertion speeds must be synchronized to ensure proper dental implant advancement. This study investigated the effects of a mismatch between rotational and linear speeds during implant insertion on primary stability. Methods Thirty identical dental implants were inserted into synthetic and natural bone at a constant rotational speed of 25 rpm. Correct insertion (CI) was defined as synchronized rotational and linear speeds. Accelerated insertion (AI) referred to a higher linear speed relative to the rotational speed. Primary stability was evaluated using maximum insertion torque (MIT), implant stability quotient (ISQ), and push-in test parameters. Results In natural bone, correct insertion produced significantly higher primary stability than accelerated insertion, as measured by MIT ( P < .001), ISQ ( P = .013), bone–implant construct stiffness ( P = .019), and maximum push-in force ( P < .001). In synthetic bone, no significant differences in stability indicators were found between insertion protocols, but axial force in the AI group was significantly higher compared to the CI group ( P = .013). For correct insertion in natural bone, a positive correlation was found between MIT and maximum push-in force (ρ = 0.867, P = .015). In contrast, no other correlation among stability indicators was statistically significant. Conclusion Mismatch between linear and rotational insertion speeds was associated with reduced primary stability in natural bone and increased axial force in synthetic bone. Clinical relevance Monitoring axial force during implant insertion, particularly in low-density bone, may provide valuable feedback on insertion accuracy, but requires further investigation.

International Dental JournalVol. 77(1)
Amirkabir University of Technology (IR)
Good health and well-being
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Mismatch Between Rotational and Linear Speed Affects Primary Stability of Dental Implants: An Experimental Study — Gholamreza Rouhi, Mahdi Farrokhi Kashtiban · International Dental Journal (2026) | TGRS Research Map | TGRS