In silico feasibility and robustness evaluation of a single-spot proton Bragg peak FLASH platform for small-animal irradiation using a generalized 3D-printed bolus
Objective FLASH radiotherapy has shown potential to reduce normal tissue toxicity while maintaining tumour control, motivating the development of clinical delivery strategies. However, most preclinical proton FLASH studies rely on transmission beams or spread out Bragg peaks, which do not reflect conformal clinical dose distributions. This work aims to develop and evaluate an irradiation setup for ultra-high dose-rate Bragg peak proton delivery to orthotopic mammary tumours in mice, employable in image-guided research rooms powered by clinical cyclotrons. Approach We built a simulation framework (Python+TOPAS) to assess a single spot, ablative configuration comprising a copper degrader, collimator, and generalised (3D-printed) bolus. Performance was investigated on a Phantom Mice Cohort (N=6) using personalised boluses and a Real Mice Cohort (N=10; µCT/PET) using a cohort-derived generalised bolus. Planning goals were D99%,GTV=19.0Gy and HIGTV≤1.30. Physical dose-rate feasibility was evaluated against a nominal 40Gy/s DRmin,skin threshold under an assumed 500nA beam current cap; this metric does not establish a biological FLASH effect. Robustness evaluation included 0.35 and 0.5mm setup errors and ±3% range uncertainties. Main results Simulated plans met target and heterogeneity goals across cohorts, with the 40Gy/s DRmin,skin criterion met under the assumed beam current conditions (Dmean,skin≈18.5-19.3Gy). The modelled dose-rate criterion was met below the 500nA beam current limit in both cohorts. Under the evaluated 0.35mm setup and ±3% range uncertainties, reduction in D99%,GTV remained within 3% for 9/10 real mouse cases. The generalised bolus achieved the simulated dosimetric performance across ten anatomically distinct mice; applicability to independent cohorts remains unestablished. Significance This in silico evaluation supports the simulated dosimetric and physical dose-rate feasibility of conformal Bragg peak proton delivery in small animals using a generalised bolus. The presented framework enables further development in image-guided proton research facilities, forming a strong basis for experimental validation and commissioning toward in situ small animal studies. .
Authors
- Stefan Both (ORCID: https://orcid.org/0000-0001-6652-6054)
- Lara Barazzuol (ORCID: https://orcid.org/0000-0002-6538-2383)
- A. Gerbershagen (ORCID: https://orcid.org/0000-0002-7648-3681)
- Jeremy P Gunawan (ORCID: https://orcid.org/0000-0002-5425-1011)
- Rutger J C de Koster (ORCID: https://orcid.org/0000-0003-3895-3969)
- Marc‐Jan van Goethem
- Rob P Coppes
Institutions
- University Medical Center Groningen (NL)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2026-11-20
- DOI
- https://doi.org/10.1088/1361-6560/aec530
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Rijksuniversiteit Groningen
- Ministerie van Economische Zaken en Klimaat