Deep learning reconstruction for low‐dose thoracic CT of calcification inserts: A phantom study of CNR, noise magnitude, and noise texture

Purpose To compare filtered back projection (FBP), 50% Adaptive Statistical Iterative Reconstruction‐Veo (ASiR), and medium‐strength deep learning image reconstruction (DLIR) for hydroxyapatite calcification inserts under different dose, tube‐voltage, and phantom‐size conditions using contrast‐to‐noise ratio (CNR), background noise magnitude, and noise power spectrum (NPS) analysis. Methods A QRM‐Thorax containing a QRM‐CCI D100 cardiac calcification insert was scanned on a GE Revolution Apex CT system using institution‐defined routine and low‐dose protocols at 80, 100, 120, and 140 kVp, with and without an Extension Ring L. Images were reconstructed at 2.5‐mm thickness using FBP, ASiR, and DLIR. CNR was evaluated for 3‐ and 5‐mm inserts at 200, 400, and 800 mg HA/cm 3 . Background noise was assessed from a calcification‐excluded mask. For NPS analysis, 10 calcification‐free images were analyzed for each acquisition condition. Total two‐dimensional NPS power, power‐weighted normalized radial NPS, and power‐weighted mean spatial frequency were calculated. Results DLIR reduced background noise magnitude and increased CNR relative to FBP across the evaluated conditions, whereas differences relative to ASiR varied with insert size, density, tube voltage, and phantom size. Across all 16 protocol, phantom‐size, and tube‐voltage combinations, ASiR and DLIR reduced total 2D NPS power by 58.1% and 75.8%, respectively, relative to FBP; DLIR reduced total power by 41.9% relative to ASiR. ASiR produced a more consistent shift toward lower power‐weighted mean spatial frequencies, averaging 12.4% below FBP, whereas the shift with DLIR was smaller and more condition dependent, averaging 3.2% below FBP. Conclusions In this controlled phantom study, DLIR reduced total noise power and improved CNR relative to FBP, but its incremental difference from ASiR was task and protocol dependent. CNR and noise magnitude should be interpreted together with NPS because nonlinear reconstruction methods can alter both noise magnitude and spatial‐frequency distribution.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-11-20
DOI
https://doi.org/10.1002/acm2.70916
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Deep learning reconstruction for low‐dose thoracic CT of calcification inserts: A phantom study of CNR, noise magnitude, and noise texture

Tou-Rong Chen, Ya‐Yun Hsiao, Hui‐Chuan Kao, Cheng‐Chang Lu
1 links
Journal of Applied Clinical Medical Physics
Radiation Dose and Imaging
article

Deep learning reconstruction for low‐dose thoracic CT of calcification inserts: A phantom study of CNR, noise magnitude, and noise texture

Tou-Rong Chen, Ya‐Yun Hsiao, Hui‐Chuan Kao, Cheng‐Chang Lu
article en

Abstract

Purpose To compare filtered back projection (FBP), 50% Adaptive Statistical Iterative Reconstruction‐Veo (ASiR), and medium‐strength deep learning image reconstruction (DLIR) for hydroxyapatite calcification inserts under different dose, tube‐voltage, and phantom‐size conditions using contrast‐to‐noise ratio (CNR), background noise magnitude, and noise power spectrum (NPS) analysis. Methods A QRM‐Thorax containing a QRM‐CCI D100 cardiac calcification insert was scanned on a GE Revolution Apex CT system using institution‐defined routine and low‐dose protocols at 80, 100, 120, and 140 kVp, with and without an Extension Ring L. Images were reconstructed at 2.5‐mm thickness using FBP, ASiR, and DLIR. CNR was evaluated for 3‐ and 5‐mm inserts at 200, 400, and 800 mg HA/cm 3 . Background noise was assessed from a calcification‐excluded mask. For NPS analysis, 10 calcification‐free images were analyzed for each acquisition condition. Total two‐dimensional NPS power, power‐weighted normalized radial NPS, and power‐weighted mean spatial frequency were calculated. Results DLIR reduced background noise magnitude and increased CNR relative to FBP across the evaluated conditions, whereas differences relative to ASiR varied with insert size, density, tube voltage, and phantom size. Across all 16 protocol, phantom‐size, and tube‐voltage combinations, ASiR and DLIR reduced total 2D NPS power by 58.1% and 75.8%, respectively, relative to FBP; DLIR reduced total power by 41.9% relative to ASiR. ASiR produced a more consistent shift toward lower power‐weighted mean spatial frequencies, averaging 12.4% below FBP, whereas the shift with DLIR was smaller and more condition dependent, averaging 3.2% below FBP. Conclusions In this controlled phantom study, DLIR reduced total noise power and improved CNR relative to FBP, but its incremental difference from ASiR was task and protocol dependent. CNR and noise magnitude should be interpreted together with NPS because nonlinear reconstruction methods can alter both noise magnitude and spatial‐frequency distribution.

Journal of Applied Clinical Medical PhysicsVol. 27(12)
National Formosa University (TW), Chung Shan Medical University Hospital (TW), Chung Shan Medical University (TW)
Good health and well-being
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon