Local Voxelwise EQD2, Prescription-Fractionation Normalisation, and Common-Fraction Equivalent Dose: A Theoretical and Computational Comparison
Equivalent dose in 2-Gy fractions (EQD2) is widely used to compare radiotherapy schedules, but different transformations of heterogeneous dose distributions need not produce interchangeable dose-volume histograms. In this in silico study, we analytically and computationally compared local voxelwise EQD2, prescription-fractionation normalisation, and a biologically effective dose-preserving common-fraction equivalent dose. Computational analyses were implemented in MATLAB R2021b using the deterministic three-dimensional phantom, with figures generated from the resulting outputs. Relative voxel doses of 0-1.2, prescription fraction sizes of 1-20 Gy, normal-tissue α/β ratios of 2-10 Gy, and target α/β ratios of 3-15 Gy were examined in a deterministic three-dimensional phantom containing two organs at risk. Prescription normalisation and local voxelwise EQD2 agree at relative doses 0 and 1; within the evaluated range their maximum absolute normalised difference occurs at a relative dose of 0.5. For 60 Gy in 20 fractions, a voxel receiving 30 Gy yielded 27.0 Gy by local voxelwise conversion, 36.0 Gy by prescription normalisation, and 33.49 Gy by common-fraction conversion. Mean doses in the adjacent organ were 11.48, 17.44, and 15.73 Gy, respectively. The three transformations preserve different mathematical properties and generate different dose-volume histograms. Consequently, physical-dose constraints should not be transferred automatically to transformed distributions without method-specific validation.
Authors
- Tae Kyu Lee (ORCID: https://orcid.org/0000-0003-0218-4942)
Institutions
- Indiana University Health (US)
Publication Details
- Journal
- Biomedical Physics & Engineering Express
- Published
- 2026-11-20
- DOI
- https://doi.org/10.1088/2057-1976/aec4e2
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00