Dynamic corneal responses from successive tonometry at different resting intervals using the Corvis ST in young myopes

Background To investigate the effects of successive measurements and resting intervals on dynamic corneal response (DCR) parameters obtained using corneal visualization Scheimpflug tonometry (Corvis ST). Methods This prospective repeated-measures observational study included young healthy myopes who underwent three successive Corvis ST measurements at four fixed intervals: 5 min (T5m), 2 min (T2m), 30 s (T30s), and immediate succession (Ti). Eleven DCR parameters were analyzed: first applanation time (A1T) and velocity (A1V), deformation amplitude (DA), highest concavity time (HCT), peak distance (PeakDist), deflection amplitude (DefleA), stiffness parameter at first applanation (SP-A1), and stress-strain index (SSI), deflection length at A1 (A1DefLen), deflection amplitude at A1 (A1DefAmp), and deflection area at A1 (A1DefArea). Linear mixed models with a random subject intercept were used to test the effects of successive measurements, intervals, and their interactions, adjusted for axial length (AL), biomechanically corrected intraocular pressure (bIOP), and central corneal thickness (CCT), with Sidak-adjusted post hoc comparisons. Results No interactions between successive measurements and intervals were detected for any parameter (all p > 0.11). Successive measurements significantly affected only DA (p = 0.001). The 2 nd measurement showed a slightly larger DA (+0.010 mm versus the first, p < 0.001). The interval significantly affected the DA (p = 0.005), HCT (p = 0.015), SP-A1 (p = 0.004), A1DefLen (p = 0.005), and A1DefAmp (p = 0.002). The largest adjusted differences were 0.01 mm (DA), 0.09 ms (HCT), 1.5 mmHg/mm (SP-A1), 0.028 mm (A1DefLen), and 0.001 mm (A1DefAmp). A longer AL was independently associated with a larger DefleA (p < 0.001) and PeakDist (p < 0.001). SSI was independent of successive measurements, intervals, bIOP, and CCT (all p > 0.11). Conclusion Corvis ST DCR parameters had only slight changes across successive measurements and different resting intervals from 5 min to no rest in young myopes. Averaging three Corvis ST measurements may be useful, regardless of the resting interval between acquisitions, although further studies are needed.

Authors

Institutions

Publication Details

Journal
Frontiers in Bioengineering and Biotechnology
Published
2026-11-20
DOI
https://doi.org/10.3389/fbioe.2026.2026392
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Dynamic corneal responses from successive tonometry at different resting intervals using the Corvis ST in young myopes

Xingyi Guo, Hongsu Wang, Han Yu Zhang, Andrew K. C. Lam et al.
1 links
Frontiers in Bioengineering and Biotechnology
Corneal surgery and disorders
article

Dynamic corneal responses from successive tonometry at different resting intervals using the Corvis ST in young myopes

Xingyi Guo, Hongsu Wang, Han Yu Zhang, Andrew K. C. Lam, Lihua Li, Byung Soo Kang, Xiaoqin Chen
article en

Abstract

Background To investigate the effects of successive measurements and resting intervals on dynamic corneal response (DCR) parameters obtained using corneal visualization Scheimpflug tonometry (Corvis ST). Methods This prospective repeated-measures observational study included young healthy myopes who underwent three successive Corvis ST measurements at four fixed intervals: 5 min (T5m), 2 min (T2m), 30 s (T30s), and immediate succession (Ti). Eleven DCR parameters were analyzed: first applanation time (A1T) and velocity (A1V), deformation amplitude (DA), highest concavity time (HCT), peak distance (PeakDist), deflection amplitude (DefleA), stiffness parameter at first applanation (SP-A1), and stress-strain index (SSI), deflection length at A1 (A1DefLen), deflection amplitude at A1 (A1DefAmp), and deflection area at A1 (A1DefArea). Linear mixed models with a random subject intercept were used to test the effects of successive measurements, intervals, and their interactions, adjusted for axial length (AL), biomechanically corrected intraocular pressure (bIOP), and central corneal thickness (CCT), with Sidak-adjusted post hoc comparisons. Results No interactions between successive measurements and intervals were detected for any parameter (all p > 0.11). Successive measurements significantly affected only DA (p = 0.001). The 2 nd measurement showed a slightly larger DA (+0.010 mm versus the first, p < 0.001). The interval significantly affected the DA (p = 0.005), HCT (p = 0.015), SP-A1 (p = 0.004), A1DefLen (p = 0.005), and A1DefAmp (p = 0.002). The largest adjusted differences were 0.01 mm (DA), 0.09 ms (HCT), 1.5 mmHg/mm (SP-A1), 0.028 mm (A1DefLen), and 0.001 mm (A1DefAmp). A longer AL was independently associated with a larger DefleA (p < 0.001) and PeakDist (p < 0.001). SSI was independent of successive measurements, intervals, bIOP, and CCT (all p > 0.11). Conclusion Corvis ST DCR parameters had only slight changes across successive measurements and different resting intervals from 5 min to no rest in young myopes. Averaging three Corvis ST measurements may be useful, regardless of the resting interval between acquisitions, although further studies are needed.

Frontiers in Bioengineering and BiotechnologyVol. 14
Hong Kong Polytechnic University (HK), Nankai University (CN), Tianjin Medical University Eye Hospital (CN)
Good health and well-being
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon