Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance

BACKGROUND: Depth of visualization (DoV) measurements are recommended as part of routine ultrasound (US) quality assurance (QA) in international guidelines. Effective QA requires measurement and analysis procedures that are repeatable, reliable, and minimally operator‑dependent. PURPOSE: This study compares three visual DoV analysis methods with semiautomated analysis tool UltraIQ to identify the approach with the highest repeatability and improved objectivity. METHODS: A total of 215 retrospective DoV measurements acquired between 2018-2022 were analyzed, including 40 transducer models from 14 US devices by six manufacturers. Visual DoV measurements were performed on the US system display and on a medical-grade display using a Picture Archiving and Communication System (PACS) using the vertical distance group and the uniformity images obtained from QA Phantom. DoV was also evaluated using SNR-based analysis in the UltraIQ software. Repeatability was evaluated using three repeated DoV measurements for 17 transducers, and reliability was assessed using the intraclass correlation coefficient (ICC). RESULTS: UltraIQ provided DoV values that were 5.0%-14.5% deeper than those obtained with visual methods. Visual DoV on the US device display was 0.7-2.0 cm shallower than UltraIQ and 0.1-0.9 cm shallower than visual DoV on PACS. Visual measurements using uniformity images on PACS were 0.4-1.1 cm shallower than UltraIQ but demonstrated the closest agreement with UltraIQ results. ICC analysis showed good to excellent reliability. For visually measured DoV, ICC values ranged 0.893-0.939, while UltraIQ provided an ICC of 0.882. CONCLUSION: Semiautomated DoV analysis using UltraIQ is as repeatable and reliable as visual assessment while providing a more objective approach for US QA.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-11-20
DOI
https://doi.org/10.1002/acm2.70883
Primary Topic
Quality and Safety in Healthcare
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance

Miia Nadhum, Antti Vehkaoja, Ullamari Hakulinen, Anna Vuorenmaa
Journal of Applied Clinical Medical Physics
Quality and Safety in Healthcare
article

Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance

Miia Nadhum, Antti Vehkaoja, Ullamari Hakulinen, Anna Vuorenmaa
article en

Abstract

BACKGROUND: Depth of visualization (DoV) measurements are recommended as part of routine ultrasound (US) quality assurance (QA) in international guidelines. Effective QA requires measurement and analysis procedures that are repeatable, reliable, and minimally operator‑dependent. PURPOSE: This study compares three visual DoV analysis methods with semiautomated analysis tool UltraIQ to identify the approach with the highest repeatability and improved objectivity. METHODS: A total of 215 retrospective DoV measurements acquired between 2018-2022 were analyzed, including 40 transducer models from 14 US devices by six manufacturers. Visual DoV measurements were performed on the US system display and on a medical-grade display using a Picture Archiving and Communication System (PACS) using the vertical distance group and the uniformity images obtained from QA Phantom. DoV was also evaluated using SNR-based analysis in the UltraIQ software. Repeatability was evaluated using three repeated DoV measurements for 17 transducers, and reliability was assessed using the intraclass correlation coefficient (ICC). RESULTS: UltraIQ provided DoV values that were 5.0%-14.5% deeper than those obtained with visual methods. Visual DoV on the US device display was 0.7-2.0 cm shallower than UltraIQ and 0.1-0.9 cm shallower than visual DoV on PACS. Visual measurements using uniformity images on PACS were 0.4-1.1 cm shallower than UltraIQ but demonstrated the closest agreement with UltraIQ results. ICC analysis showed good to excellent reliability. For visually measured DoV, ICC values ranged 0.893-0.939, while UltraIQ provided an ICC of 0.882. CONCLUSION: Semiautomated DoV analysis using UltraIQ is as repeatable and reliable as visual assessment while providing a more objective approach for US QA.

Journal of Applied Clinical Medical PhysicsVol. 27(12)
Tampere University (FI), Tampere University Hospital (FI), University of Tampere (FI)
Tays
Good health and well-being
Quality and Safety in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon