Trans-epithelial/endothelial electrical resistance (TEER) measurements are a key nondestructive measurement of the barrier integrity of cells grown on inserts. To support the use of TEER measurements in bioassays, we have evaluated the sources of variability in a case study involving a chamber-based TEER system to assess a gut co-culture model. Data from 23 separate experiments, each lasting 22 d, were evaluated. Multiple lines of evidence suggested minimal variability within a run, a period during which several plates were analyzed sequentially. In contrast, run-to-run variability was substantially larger, even when two runs were performed several hours apart on the same day. Bayesian analysis of both KCl and blank insert data revealed that run-to-run variability was among the largest sources of variability, with relative uncertainty values of 3 % to 6 % of the mean value. Correlational analysis investigated whether deviations of KCl or blank insert TEER values within a run compared to long-term median values were correlated with deviations of the biculture gut model values within that same run from their long-term median values. Unexpectedly, the KCl values were not correlated, while the blank insert values had a weak correlation. We also evaluated the information gained from different control measurements. The findings from this study can influence how statistical analyses are performed on TEER results and can inform the design of robust TEER methods. For example, the approaches described here can be used to characterize variability within and among runs in TEER measurements for other methods.

Authors

Institutions

Publication Details

Journal
bioRxiv (Cold Spring Harbor Laboratory)
Published
2026-11-19
DOI
https://doi.org/10.64898/2026.11.06.764605
Primary Topic
Animal testing and alternatives
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Evaluation of sources of variability in a TEER (trans-epithelial/endothelial electrical resistance) system

Steven Lund, Robert Gutiérrez, Elijah Joel Petersen, Vytas Reipa
bioRxiv (Cold Spring Harbor Laboratory)
Animal testing and alternatives
preprint

Evaluation of sources of variability in a TEER (trans-epithelial/endothelial electrical resistance) system

Steven Lund, Robert Gutiérrez, Elijah Joel Petersen, Vytas Reipa
preprint en

Abstract

Trans-epithelial/endothelial electrical resistance (TEER) measurements are a key nondestructive measurement of the barrier integrity of cells grown on inserts. To support the use of TEER measurements in bioassays, we have evaluated the sources of variability in a case study involving a chamber-based TEER system to assess a gut co-culture model. Data from 23 separate experiments, each lasting 22 d, were evaluated. Multiple lines of evidence suggested minimal variability within a run, a period during which several plates were analyzed sequentially. In contrast, run-to-run variability was substantially larger, even when two runs were performed several hours apart on the same day. Bayesian analysis of both KCl and blank insert data revealed that run-to-run variability was among the largest sources of variability, with relative uncertainty values of 3 % to 6 % of the mean value. Correlational analysis investigated whether deviations of KCl or blank insert TEER values within a run compared to long-term median values were correlated with deviations of the biculture gut model values within that same run from their long-term median values. Unexpectedly, the KCl values were not correlated, while the blank insert values had a weak correlation. We also evaluated the information gained from different control measurements. The findings from this study can influence how statistical analyses are performed on TEER results and can inform the design of robust TEER methods. For example, the approaches described here can be used to characterize variability within and among runs in TEER measurements for other methods.

bioRxiv (Cold Spring Harbor Laboratory)
National Institute of Standards and Technology (US)
Animal testing and alternatives
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon