Digital Thoracic Drainage System-Derived Circuit Pressure and Air-Flow Changes in Relation to Echocardiographic Indices of Cardiac Function after Chest Tube Placement

OBJECTIVE Chest tube drainage restores negative intrathoracic pressure, which might alter cardiac loading. Digital thoracic drainage systems continuously record pressure and airflow. We examined device-derived pressure and air-flow trends over 48 hours and echocardiographic changes between 24 and 48 hours. MATERIAL AND METHODS Forty-one chest tube drainage episodes in 39 adult, non-mechanically ventilated patients were studied prospectively. Device records were analysed at 10-minute intervals, providing preset suction level, measured circuit pressure, and air flow. Echocardiography was performed at 24 and 48 hours. Continuous time-series analyses included 29 episodes, 28 of which were evaluable within the predefined 48-hour window. False discovery rate correction was applied to multiple comparisons. RESULTS The preset suction level remained unchanged during the first 48 hours in 21 of 29 episodes (72%), and the measured circuit pressure was closely related to it at each time point (R 2 = 0.838-0.955). Air leak decreased from a median of 61.6 mL/min at baseline to 2.3 mL/min at 24 hours ( P < 0.001), with no further change at 48 hours (1.3 mL/min; P = 0.55). After correction, ejection fraction (q = 0.011) and mitral E velocity (q = 0.007) were higher at 48 hours. No device-derived variable was associated with echocardiographic changes after correction for multiple comparisons. CONCLUSION Measured circuit pressure appeared to reflect regulated drainage-system pressure more closely than pleural physiology. Most reduction in air leak occurred before the first echocardiographic assessment, potentially limiting the detection of cardiopulmonary associations. These findings support a cautious interpretation of digital drainage parameters and may help define the assessment window for future studies.

Authors

Institutions

Publication Details

Journal
Thoracic research and practice
Published
2026-11-19
DOI
https://doi.org/10.4274/thoracrespract.2026.2026-8-3
Primary Topic
Pleural and Pulmonary Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Digital Thoracic Drainage System-Derived Circuit Pressure and Air-Flow Changes in Relation to Echocardiographic Indices of Cardiac Function after Chest Tube Placement

Atilla Can, Abdullah İçli, Mehmet Burhan Oflaz, Tamer Altınok
Thoracic research and practice
Pleural and Pulmonary Diseases
article

Digital Thoracic Drainage System-Derived Circuit Pressure and Air-Flow Changes in Relation to Echocardiographic Indices of Cardiac Function after Chest Tube Placement

Atilla Can, Abdullah İçli, Mehmet Burhan Oflaz, Tamer Altınok
article en

Abstract

OBJECTIVE Chest tube drainage restores negative intrathoracic pressure, which might alter cardiac loading. Digital thoracic drainage systems continuously record pressure and airflow. We examined device-derived pressure and air-flow trends over 48 hours and echocardiographic changes between 24 and 48 hours. MATERIAL AND METHODS Forty-one chest tube drainage episodes in 39 adult, non-mechanically ventilated patients were studied prospectively. Device records were analysed at 10-minute intervals, providing preset suction level, measured circuit pressure, and air flow. Echocardiography was performed at 24 and 48 hours. Continuous time-series analyses included 29 episodes, 28 of which were evaluable within the predefined 48-hour window. False discovery rate correction was applied to multiple comparisons. RESULTS The preset suction level remained unchanged during the first 48 hours in 21 of 29 episodes (72%), and the measured circuit pressure was closely related to it at each time point (R 2 = 0.838-0.955). Air leak decreased from a median of 61.6 mL/min at baseline to 2.3 mL/min at 24 hours ( P < 0.001), with no further change at 48 hours (1.3 mL/min; P = 0.55). After correction, ejection fraction (q = 0.011) and mitral E velocity (q = 0.007) were higher at 48 hours. No device-derived variable was associated with echocardiographic changes after correction for multiple comparisons. CONCLUSION Measured circuit pressure appeared to reflect regulated drainage-system pressure more closely than pleural physiology. Most reduction in air leak occurred before the first echocardiographic assessment, potentially limiting the detection of cardiopulmonary associations. These findings support a cautious interpretation of digital drainage parameters and may help define the assessment window for future studies.

Thoracic research and practice
Selçuk University (TR), Necmettin Erbakan University (TR)
Pleural and Pulmonary Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon