Our experience in developing an all-ceramic first carpal-metacarpal joint endoprosthesis

Introduction. Risarthrosis is one of the common causes of disability in patients. Traditional interventions - trapezectomy and arthrodesis - have a number of serious complications. Therefore, the development of alternative surgical methods is relevant, including the arthroplasty of the first carpal-metacarpal joint (CMC-1). The aim of the work was to develop and experimentally substantiate an individual two-component unrelated whole-ceramic endoprosthesis of the CMC-1 for the surgical treatment of risarthrosis. Methods. For this study, cadaveric material was used: 5 male and 7 female hands, a total of 12 operations were performed. At the Research Institute of Bionics and Personalized Medicine, SamSMU designed and developed a two-component all-ceramic CMC-1 endoprosthesis, as well as special tools for installing implant components. Results. Based on computed tomography data, the design of an individual endoprosthesis was simulated. Further, computer modeling was carried out to create an individual virtual model of the implant. The distal and proximal components of the zirconium ceramic implant were manufactured by machining on numerically controlled machines. The proximal component was positioned on the articular surface of the trapezius bone, the distal component was mounted with an impactor and soft hammering. After installing the components of the endoprosthesis, load tests were CMC-1 performed, while the failure of the implant design due to its proximal component was revealed. Discussion. Currently, none of the surgical treatments for risarthrosis have clear benefits. To continue research for the development of an endoprosthesis, it is necessary to rethink the geometry of the implant leg, introduce additional anchor fixation, test various coatings to increase the coefficient of friction and osteointegration. Conclusion. An individual two-component unbound whole-ceramic design of the CMC-1 endoprosthesis for the surgical treatment of risarthrosis was experimentally substantiated. Despite the successful completion of the design and manufacturing stages, the design turned out to be mechanically untenable due to the ineffective fixation of the proximal component. significant experience was gained in the course of work, which can be used to develop new design options.

Authors

Institutions

Publication Details

Journal
Aspirantskiy Vestnik Povolzhiya
Published
2026-11-19
DOI
https://doi.org/10.35693/avp704102
Primary Topic
Orthopedic Surgery and Rehabilitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Our experience in developing an all-ceramic first carpal-metacarpal joint endoprosthesis

Gennadiy P. Kotelnikov, Andrey N. Nikolaenko, S. O. Doroganov, A. P. Borisov
Aspirantskiy Vestnik Povolzhiya
Orthopedic Surgery and Rehabilitation
article

Our experience in developing an all-ceramic first carpal-metacarpal joint endoprosthesis

Gennadiy P. Kotelnikov, Andrey N. Nikolaenko, S. O. Doroganov, A. P. Borisov
article en

Abstract

Introduction. Risarthrosis is one of the common causes of disability in patients. Traditional interventions - trapezectomy and arthrodesis - have a number of serious complications. Therefore, the development of alternative surgical methods is relevant, including the arthroplasty of the first carpal-metacarpal joint (CMC-1). The aim of the work was to develop and experimentally substantiate an individual two-component unrelated whole-ceramic endoprosthesis of the CMC-1 for the surgical treatment of risarthrosis. Methods. For this study, cadaveric material was used: 5 male and 7 female hands, a total of 12 operations were performed. At the Research Institute of Bionics and Personalized Medicine, SamSMU designed and developed a two-component all-ceramic CMC-1 endoprosthesis, as well as special tools for installing implant components. Results. Based on computed tomography data, the design of an individual endoprosthesis was simulated. Further, computer modeling was carried out to create an individual virtual model of the implant. The distal and proximal components of the zirconium ceramic implant were manufactured by machining on numerically controlled machines. The proximal component was positioned on the articular surface of the trapezius bone, the distal component was mounted with an impactor and soft hammering. After installing the components of the endoprosthesis, load tests were CMC-1 performed, while the failure of the implant design due to its proximal component was revealed. Discussion. Currently, none of the surgical treatments for risarthrosis have clear benefits. To continue research for the development of an endoprosthesis, it is necessary to rethink the geometry of the implant leg, introduce additional anchor fixation, test various coatings to increase the coefficient of friction and osteointegration. Conclusion. An individual two-component unbound whole-ceramic design of the CMC-1 endoprosthesis for the surgical treatment of risarthrosis was experimentally substantiated. Despite the successful completion of the design and manufacturing stages, the design turned out to be mechanically untenable due to the ineffective fixation of the proximal component. significant experience was gained in the course of work, which can be used to develop new design options.

Aspirantskiy Vestnik Povolzhiya
Samara State Medical University (RU)
Orthopedic Surgery and Rehabilitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon