Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection

Background: Artificial intelligence (AI) can detect hydronephrosis and estimate prostate volume on ultrasound, but most systems rely on expert-acquired images from conventional console ultrasound systems. We developed an on-device AI model for handheld point-of-care ultrasound (POCUS) and prospectively evaluated it on nurse-acquired scans for bedside triage/decision support. Methods: The model was trained using retrospective console images (101 renal and 100 prostate cases) and prospective POCUS development clips (44 renal and 88 prostate patients), then evaluated prospectively on ward nurse-acquired renal and suprapubic POCUS videos after brief instruction. Reference standards were blinded expert adjudication of hydronephrosis and expert planimetry for prostate volume. Primary outcomes were sensitivity/specificity for hydronephrosis detection and screening performance for severe prostate enlargement (≥50 mL). Results: For hydronephrosis detection, 32 evaluable patients were analyzed (one kidney per patient; prevalence 46.9%). The AI achieved a sensitivity of 93.3% (95% CI 70.2–98.8) and a specificity of 82.4% (59.0–93.8). For severe prostate enlargement (≥50 mL; n = 28), sensitivity and specificity were both 71.4%. Agreement for continuous prostate volume was moderate: ICC(A,1) = 0.789; Bland–Altman bias +5.1 mL, with 95% limits −25.0 to +35.1 mL. Conclusions: On nurse-acquired handheld POCUS, AI showed high sensitivity for hydronephrosis and moderate agreement for prostate-volume estimation, supporting its feasibility for triage support while requiring further validation.

Authors

Institutions

Publication Details

Journal
POCUS Journal
Published
2026-11-18
DOI
https://doi.org/10.24908/pocusj.v11i02.20395
Citations
1
Primary Topic
Ultrasound in Clinical Applications
Type
article
Field-Weighted Citation Impact
2.59
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection

Masaru Matsumoto, Motofumi Suzuki, Atsuo Kawamoto, Yuka Miura et al.
1 citations
POCUS Journal
Ultrasound in Clinical Applications
2.59
article

Nurse-Validated Handheld POCUS Artificial Intelligence (AI) for Automated Prostate Volume Estimation and Hydronephrosis Detection

Masaru Matsumoto, Motofumi Suzuki, Atsuo Kawamoto, Yuka Miura, Nao Tamai, Mikihiko Karube, Sanae Wakita, Masaya Sato, Hiromi Sanada, Masuda Minoru, Tamaki Kobayashi, Hiroaki Gotoh
article en
1 citations

Abstract

Background: Artificial intelligence (AI) can detect hydronephrosis and estimate prostate volume on ultrasound, but most systems rely on expert-acquired images from conventional console ultrasound systems. We developed an on-device AI model for handheld point-of-care ultrasound (POCUS) and prospectively evaluated it on nurse-acquired scans for bedside triage/decision support. Methods: The model was trained using retrospective console images (101 renal and 100 prostate cases) and prospective POCUS development clips (44 renal and 88 prostate patients), then evaluated prospectively on ward nurse-acquired renal and suprapubic POCUS videos after brief instruction. Reference standards were blinded expert adjudication of hydronephrosis and expert planimetry for prostate volume. Primary outcomes were sensitivity/specificity for hydronephrosis detection and screening performance for severe prostate enlargement (≥50 mL). Results: For hydronephrosis detection, 32 evaluable patients were analyzed (one kidney per patient; prevalence 46.9%). The AI achieved a sensitivity of 93.3% (95% CI 70.2–98.8) and a specificity of 82.4% (59.0–93.8). For severe prostate enlargement (≥50 mL; n = 28), sensitivity and specificity were both 71.4%. Agreement for continuous prostate volume was moderate: ICC(A,1) = 0.789; Bland–Altman bias +5.1 mL, with 95% limits −25.0 to +35.1 mL. Conclusions: On nurse-acquired handheld POCUS, AI showed high sensitivity for hydronephrosis and moderate agreement for prostate-volume estimation, supporting its feasibility for triage support while requiring further validation.

POCUS JournalVol. 11(02)
Fujita Health University (JP), Tokyo University of Science (JP), Ishikawa Prefectural Nursing University (JP), Tokyo Medical University Hospital (JP), University of Tokyo Hospital (JP), Mishima General Hospital (JP), Global Nursing Research Center, The University of Tokyo (JP), The University of Tokyo (JP), Yokohama City University (JP)
Openalex Percentile: Top 17%
Ultrasound in Clinical Applications
2.59
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon