A short embedding protocol for diagnostic electron microscopy: long-term implementation and diagnostic outcomes in a high-volume clinical series

Epoxy resin polymerization is a lengthy step in tissue preparation for transmission electron microscopy (TEM). We evaluated whether reducing the polymerization step alone to 40 minutes at 120 °C, with upstream processing unchanged, could maintain diagnostic quality. The protocol has been the sole routine method in a high-volume, accredited clinical TEM laboratory for more than 10 years. Among more than 30,000 consecutive cases, no embedding-related diagnostic inadequacy was identified in consultant pathologist sign-out records. In a prospective, blinded, split-specimen comparison of renal biopsies, both pathologists scored 4 or 5 on every feature for both methods. Nearly all scores were at the upper end of the scale, indicating comparable performance but not establishing statistical equivalence. Glomerular basement membrane thickness was closely similar (mean difference -12.7 nm, ~4%), within measurement variability and below diagnostic detection thresholds. Because curing behavior depends on the resin formulation and oven conditions, local validation is required before implementation.

Authors

Institutions

Publication Details

Journal
Ultrastructural Pathology
Published
2026-11-18
DOI
https://doi.org/10.1080/01913123.2026.2739247
Primary Topic
Renal Diseases and Glomerulopathies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A short embedding protocol for diagnostic electron microscopy: long-term implementation and diagnostic outcomes in a high-volume clinical series

Xin Zhou, Hui Zhou
Ultrastructural Pathology
Renal Diseases and Glomerulopathies
article

A short embedding protocol for diagnostic electron microscopy: long-term implementation and diagnostic outcomes in a high-volume clinical series

Xin Zhou, Hui Zhou
article en

Abstract

Epoxy resin polymerization is a lengthy step in tissue preparation for transmission electron microscopy (TEM). We evaluated whether reducing the polymerization step alone to 40 minutes at 120 °C, with upstream processing unchanged, could maintain diagnostic quality. The protocol has been the sole routine method in a high-volume, accredited clinical TEM laboratory for more than 10 years. Among more than 30,000 consecutive cases, no embedding-related diagnostic inadequacy was identified in consultant pathologist sign-out records. In a prospective, blinded, split-specimen comparison of renal biopsies, both pathologists scored 4 or 5 on every feature for both methods. Nearly all scores were at the upper end of the scale, indicating comparable performance but not establishing statistical equivalence. Glomerular basement membrane thickness was closely similar (mean difference -12.7 nm, ~4%), within measurement variability and below diagnostic detection thresholds. Because curing behavior depends on the resin formulation and oven conditions, local validation is required before implementation.

Ultrastructural Pathology
Baylor University Medical Center (US), Quest Diagnostics (United States) (US)
Renal Diseases and Glomerulopathies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon