In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement

Continuous glucose monitoring (CGM) sensors are evaluated almost exclusively in clinical studies. We investigated the dynamic response of one factory-calibrated CGM system on a programmable in vitro flow platform, with glucose measured independently by a SUPER GL2 analyzer as comparator. Twelve sensors were studied: four during verification of the measurement procedure (Stage 1) and eight tested on two consecutive days (Stage 2) under a 90 → 270 → 90 mg/dL profile. Amplitude transfer (KA) and apparent rising- and falling-phase slope transfer (Kup, Kdown) were calculated without lag correction. Stage 1 results were exploratory owing to an unexplained CGM displacement of approximately 35 min. The delivered profile differed from the programmed one in every Stage 2 run (upper plateaus 254–268 vs. 270 mg/dL). Mean KA was 0.731 and 0.731, Kup 0.749 and 0.746, and Kdown 0.729 and 0.719 on the two days; between sensors, KA ranged from 0.61–0.84. Near-zero mean paired changes reflected opposite group shifts (KA +0.026 and −0.025). Descriptive ICC(A,1) values of 0.919–0.952 (seven or eight pairs from two shared runs) indicate preserved between-sensor differences rather than established reproducibility. Bias (≈−50 mg/dL) and MARD (31–32%) apply only to these in vitro conditions; the findings do not establish clinical performance or long-term stability.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-11-19
DOI
https://doi.org/10.3390/s26227297
Primary Topic
Diabetes Management and Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement

Момыналиев Куват Темиргалиевич, Elena V. Khoroshun, Vladimir A Kozlov, Igor V. Ivanov
Sensors
Diabetes Management and Research
article

In Vitro Dynamic Characterization of a Continuous Glucose Monitoring System Using an Independently Measured Glucose Profile: Amplitude Transfer, Rate Transfer, and Consecutive-Day Agreement

Момыналиев Куват Темиргалиевич, Elena V. Khoroshun, Vladimir A Kozlov, Igor V. Ivanov
article en

Abstract

Continuous glucose monitoring (CGM) sensors are evaluated almost exclusively in clinical studies. We investigated the dynamic response of one factory-calibrated CGM system on a programmable in vitro flow platform, with glucose measured independently by a SUPER GL2 analyzer as comparator. Twelve sensors were studied: four during verification of the measurement procedure (Stage 1) and eight tested on two consecutive days (Stage 2) under a 90 → 270 → 90 mg/dL profile. Amplitude transfer (KA) and apparent rising- and falling-phase slope transfer (Kup, Kdown) were calculated without lag correction. Stage 1 results were exploratory owing to an unexplained CGM displacement of approximately 35 min. The delivered profile differed from the programmed one in every Stage 2 run (upper plateaus 254–268 vs. 270 mg/dL). Mean KA was 0.731 and 0.731, Kup 0.749 and 0.746, and Kdown 0.729 and 0.719 on the two days; between sensors, KA ranged from 0.61–0.84. Near-zero mean paired changes reflected opposite group shifts (KA +0.026 and −0.025). Descriptive ICC(A,1) values of 0.919–0.952 (seven or eight pairs from two shared runs) indicate preserved between-sensor differences rather than established reproducibility. Bias (≈−50 mg/dL) and MARD (31–32%) apply only to these in vitro conditions; the findings do not establish clinical performance or long-term stability.

SensorsVol. 26(22)
Federal Agency for Health and Social Development (RU)
Diabetes Management and Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon