Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen

Two Fabry–Pérot (FP) refractometers for the realization of the pascal, a novel single-cavity system and a well-established dual-cavity system, are independently characterized and compared. Both systems operate with N2 at pressures up to 100 kPa and are evaluated using molecular data that provide traceability to helium. The two systems yield comparable expanded uncertainties (k = 2), given by [(1.6 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the single-cavity system and [(1.8 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the dual-cavity system, demonstrating that the novel single-cavity architecture can achieve a pressure realization performance essentially equivalent to that of the established dual-cavity system. For both systems, the uncertainty is dominated by the uncertainties associated with the molar refractivity and second density virial coefficient of N2. A direct comparison of the two systems using the diffusion-isolated pressure equilibration methodology shows excellent agreement, with systematic linear and quadratic response differences of 0.17 × 10−6p and 9.2 × 10−12p2/Pa, respectively. Over an 18-day measurement campaign, the differential assessments exhibit a repeatability of 0.6 × 10−6 (k = 2), with virtually no detectable drift. These results demonstrate consistent performance between two independently realized refractometers based on slightly different cavity architectures, supporting the robustness of FP-based refractometry for high-accuracy pressure realization. They also highlight the need for improved determinations of molecular parameters of N2 that are traceable to helium, particularly its molar refractivity and second density virial coefficient.

Authors

Institutions

Publication Details

Journal
Journal of Vacuum Science & Technology B Nanotechnology and Microelectronics Materials Processing Measurement and Phenomena
Published
2026-11-19
DOI
https://doi.org/10.1116/6.0005912
Primary Topic
Scientific Measurement and Uncertainty Evaluation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen

Isak Silander, Martin Zelán, Johan Zakrisson, Ove Axner
Journal of Vacuum Science & Technology B Nanotechnology and Microelectronics Materials Processing Measurement and Phenomena
Scientific Measurement and Uncertainty Evaluation
article

Characterization and comparison of two independent Fabry–Pérot refractometers for the realization of the pascal in nitrogen

Isak Silander, Martin Zelán, Johan Zakrisson, Ove Axner
article en

Abstract

Two Fabry–Pérot (FP) refractometers for the realization of the pascal, a novel single-cavity system and a well-established dual-cavity system, are independently characterized and compared. Both systems operate with N2 at pressures up to 100 kPa and are evaluated using molecular data that provide traceability to helium. The two systems yield comparable expanded uncertainties (k = 2), given by [(1.6 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the single-cavity system and [(1.8 mPa)2 + (13 × 10−6p)2 + (37 × 10−12p2/Pa)2]1/2 for the dual-cavity system, demonstrating that the novel single-cavity architecture can achieve a pressure realization performance essentially equivalent to that of the established dual-cavity system. For both systems, the uncertainty is dominated by the uncertainties associated with the molar refractivity and second density virial coefficient of N2. A direct comparison of the two systems using the diffusion-isolated pressure equilibration methodology shows excellent agreement, with systematic linear and quadratic response differences of 0.17 × 10−6p and 9.2 × 10−12p2/Pa, respectively. Over an 18-day measurement campaign, the differential assessments exhibit a repeatability of 0.6 × 10−6 (k = 2), with virtually no detectable drift. These results demonstrate consistent performance between two independently realized refractometers based on slightly different cavity architectures, supporting the robustness of FP-based refractometry for high-accuracy pressure realization. They also highlight the need for improved determinations of molecular parameters of N2 that are traceable to helium, particularly its molar refractivity and second density virial coefficient.

Journal of Vacuum Science & Technology B Nanotechnology and Microelectronics Materials Processing Measurement and PhenomenaVol. 44(6)
RISE Research Institutes of Sweden (SE), Umeå University (SE)
Scientific Measurement and Uncertainty Evaluation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon