Test–retest repeatability of MBSS-T1: a model-based, subject-specific self-supervised motion correction method for robust cardiac T1 mapping

Objective. Reliable myocardial T1 mapping is essential for quantitative cardiac tissue characterization but remains sensitive to respiratory motion, particularly during free-breathing acquisitions. This study evaluated the test–retest repeatability of MBSS-T1, a subject-specific, model-based, self-supervised deep-learning method for motion-robust myocardial T1 mapping. Approach. Seventeen healthy male volunteers aged 20–36 years underwent 3 T cardiac MRI using a 5(3)3 MOLLI acquisition under breath-hold (BH) and free-breathing (FB) conditions. Same-day repeat scans were acquired after a 20 min interval that included removal from and repositioning in the scanner. T1 maps were reconstructed using (1) MBSS-T1, (2) the contrast-agnostic deep-learning registration method SynthMorph, (3) scanner-inline motion correction, and (4) no motion correction. Repeatability was evaluated using intraclass correlation coefficients (ICC), relative Bland–Altman analysis, within-subject coefficients of variation (CVw), relative 95% repeatability coefficients (RC95), and variance-based bootstrap inference with Holm correction. Main results. All four methods included 17 complete participant-level test–retest pairs under each breathing condition. Under BH conditions, the global ICC point estimates were 0.77, 0.64, 0.65, and 0.60 for MBSS-T1, SynthMorph, scanner-inline correction, and no correction, respectively. Under FB conditions, the corresponding values were 0.41, 0.32, 0.27, and 0.16. MBSS-T1 also had the lowest descriptive CVw under both conditions. However, none of the paired comparisons of absolute test–retest differences was significant after Holm correction (all adjusted p ≥ 0.623). The smallest Holm-adjusted variance-ratio result was obtained for MBSS-T1 versus no correction under FB conditions (variance ratio 0.43, 95% CI 0.30–0.68; adjusted p = 0.058), which did not meet the prespecified significance threshold. Significance. MBSS-T1 yielded favourable descriptive repeatability estimates, particularly under FB conditions, but no pairwise comparison remained statistically significant after correction for multiple comparisons. These results characterize same-day, same-scanner measurement precision rather than absolute T1 accuracy and require confirmation in larger and more clinically diverse cohorts.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-19
DOI
https://doi.org/10.1088/1361-6560/aec4b5
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Test–retest repeatability of MBSS-T1: a model-based, subject-specific self-supervised motion correction method for robust cardiac T1 mapping

Daphna Link‐Sourani, Moti Freiman, Eyal Hanania, Ilya Volovik et al.
Physics in Medicine and Biology
Advanced MRI Techniques and Applications
article

Test–retest repeatability of MBSS-T1: a model-based, subject-specific self-supervised motion correction method for robust cardiac T1 mapping

Daphna Link‐Sourani, Moti Freiman, Eyal Hanania, Ilya Volovik, Adi Zehavi
article en

Abstract

Objective. Reliable myocardial T1 mapping is essential for quantitative cardiac tissue characterization but remains sensitive to respiratory motion, particularly during free-breathing acquisitions. This study evaluated the test–retest repeatability of MBSS-T1, a subject-specific, model-based, self-supervised deep-learning method for motion-robust myocardial T1 mapping. Approach. Seventeen healthy male volunteers aged 20–36 years underwent 3 T cardiac MRI using a 5(3)3 MOLLI acquisition under breath-hold (BH) and free-breathing (FB) conditions. Same-day repeat scans were acquired after a 20 min interval that included removal from and repositioning in the scanner. T1 maps were reconstructed using (1) MBSS-T1, (2) the contrast-agnostic deep-learning registration method SynthMorph, (3) scanner-inline motion correction, and (4) no motion correction. Repeatability was evaluated using intraclass correlation coefficients (ICC), relative Bland–Altman analysis, within-subject coefficients of variation (CVw), relative 95% repeatability coefficients (RC95), and variance-based bootstrap inference with Holm correction. Main results. All four methods included 17 complete participant-level test–retest pairs under each breathing condition. Under BH conditions, the global ICC point estimates were 0.77, 0.64, 0.65, and 0.60 for MBSS-T1, SynthMorph, scanner-inline correction, and no correction, respectively. Under FB conditions, the corresponding values were 0.41, 0.32, 0.27, and 0.16. MBSS-T1 also had the lowest descriptive CVw under both conditions. However, none of the paired comparisons of absolute test–retest differences was significant after Holm correction (all adjusted p ≥ 0.623). The smallest Holm-adjusted variance-ratio result was obtained for MBSS-T1 versus no correction under FB conditions (variance ratio 0.43, 95% CI 0.30–0.68; adjusted p = 0.058), which did not meet the prespecified significance threshold. Significance. MBSS-T1 yielded favourable descriptive repeatability estimates, particularly under FB conditions, but no pairwise comparison remained statistically significant after correction for multiple comparisons. These results characterize same-day, same-scanner measurement precision rather than absolute T1 accuracy and require confirmation in larger and more clinically diverse cohorts.

Physics in Medicine and Biology
Technion – Israel Institute of Technology (IL), Bnai Zion Medical Center (IL)
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon