The numerical study on cardiac ablation considering dynamic blood flow by high frequency bipolar pulses with determined lethal threshold in vitro

Objective: Pulsed Field Ablation (PFA) is a promising non-thermal energy modality for catheter-based cardiac ablation. Treatment planning using numerical simulations of PFA is important for predicting the ablation region. However, owing to the high voltage-to-electrode distance ratio, thermal effects should be considered during treatment planning. This study aims to analyse the ablation region for cardiac treatment planning considering the thermal effect. Approach: this study aimed first to determine the lethal electric field threshold (LEFT) of H9C2 myocardial cells exposed to high-frequency bipolar pulses (HFBPs). Then two simplified pulmonary vein (PV) models with a circular electrode array were constructed to investigate electric field and thermal distributions, accounting for dynamic blood flow and conductivity variation. Main results: cell experimental results showed that under the condition of equivalent energized time, fewer pulses per burst and prolonged interpulse delay of HFBPs can reduce the LEFT of myocardial cells. The LET reached 720 V/cm when applying 120 HFBPs of 1000 μs duration. Numerical simulations revealed that temperature rises sharply when blood flow is neglected, with a maximum temperature of 67 ℃. However, thermal effects are substantially limited (peak temperature of 38.3 ℃) due to blood flow, even when HFBPs with a high voltage-to-electrode distance (1250 V) are applied. The temperature rise rapidly declines and reaches a plateau at higher blood velocities. HFBPs can produce a full circumferential ablation region, and the ablation width increases linearly with rising applied voltage. Significance: this study demonstrates that dynamic blood flow and conductivity effects should be considered for myocardial ablation assessments, which may assist clinicians in precise pulse selection.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-19
DOI
https://doi.org/10.1088/1361-6560/aec497
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

The numerical study on cardiac ablation considering dynamic blood flow by high frequency bipolar pulses with determined lethal threshold in vitro

Yanpeng Lv, Gangqiong Liu, Yawen Tian, Lin Liu et al.
Physics in Medicine and Biology
Cardiac Arrhythmias and Treatments
article

The numerical study on cardiac ablation considering dynamic blood flow by high frequency bipolar pulses with determined lethal threshold in vitro

Yanpeng Lv, Gangqiong Liu, Yawen Tian, Lin Liu, Yiqing Lu
article en

Abstract

Objective: Pulsed Field Ablation (PFA) is a promising non-thermal energy modality for catheter-based cardiac ablation. Treatment planning using numerical simulations of PFA is important for predicting the ablation region. However, owing to the high voltage-to-electrode distance ratio, thermal effects should be considered during treatment planning. This study aims to analyse the ablation region for cardiac treatment planning considering the thermal effect. Approach: this study aimed first to determine the lethal electric field threshold (LEFT) of H9C2 myocardial cells exposed to high-frequency bipolar pulses (HFBPs). Then two simplified pulmonary vein (PV) models with a circular electrode array were constructed to investigate electric field and thermal distributions, accounting for dynamic blood flow and conductivity variation. Main results: cell experimental results showed that under the condition of equivalent energized time, fewer pulses per burst and prolonged interpulse delay of HFBPs can reduce the LEFT of myocardial cells. The LET reached 720 V/cm when applying 120 HFBPs of 1000 μs duration. Numerical simulations revealed that temperature rises sharply when blood flow is neglected, with a maximum temperature of 67 ℃. However, thermal effects are substantially limited (peak temperature of 38.3 ℃) due to blood flow, even when HFBPs with a high voltage-to-electrode distance (1250 V) are applied. The temperature rise rapidly declines and reaches a plateau at higher blood velocities. HFBPs can produce a full circumferential ablation region, and the ablation width increases linearly with rising applied voltage. Significance: this study demonstrates that dynamic blood flow and conductivity effects should be considered for myocardial ablation assessments, which may assist clinicians in precise pulse selection.

Physics in Medicine and Biology
Zhengzhou University (CN), First Affiliated Hospital of Zhengzhou University (CN), Henan University of Animal Husbandry and Economy (CN)
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon