RadCalc 3D Monte-Carlo: tuning and independent dosimetric validation of radiotherapy treatment plans

Objective. This work aimed at implementing a new recalculation software, ‘RadCalc’ (LAP GmbH Laser Applikationen), independent of the monitor units (MUs) and dosimetric verification of the radiotherapy treatment plans specific for each patient, using the Monte-Carlo (MC) algorithm. Approach. The experimental procedure for the simulation of the treatment plans by the software was started by measuring dosimetric parameters of the operating machine (Elekta Versa HD linear accelerator), subsequently imported into the software. After that, the software tuning on the treatment planning system Monaco v. 5.53 (Elekta) was carried out, which allowed to extract the most suitable Additional Radiation to Light Field Offset parameter to obtain the best performing MC algorithm model for the simulation of the treatment plans. Finally, the above-mentioned model was validated through a statistical analysis comparing the plans recalculated by RadCalc and those experimentally verified on the OCTAVIUS 4D phantom (PTW) and Octavius 1500 (PTW) Detector. Main results. The obtained results showed a good statistical agreement both in the tuning and in the validation part of the MC algorithm model used. The effectiveness of using the RadCalc software was further demonstrated by the high accuracy of the recalculation of the MUs. Significance. It should be noted that the secondary recalculation of the dose, in accordance with Italian legislation on patient radiation protection is one of the actions that the medical physics specialist must ‘prepare for the evaluation and correct verification of the doses administered to patients’ . In this context, it has to be emphasized that a multicenter study group is being set up, within the Italian Association of Medical and Health Physics (AIFM), which brings together the users of RadCalc, aimed at exchanging experiences and improving the performance of the aforementioned software.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-19
DOI
https://doi.org/10.1088/2057-1976/aec16e
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

RadCalc 3D Monte-Carlo: tuning and independent dosimetric validation of radiotherapy treatment plans

F. Caridi, Domenico Majolino, Valentina Venuti, Giuseppe Paladini et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

RadCalc 3D Monte-Carlo: tuning and independent dosimetric validation of radiotherapy treatment plans

F. Caridi, Domenico Majolino, Valentina Venuti, Giuseppe Paladini, Salvatore Pittera, B.M. Itri, Giuseppe Sceni, Jacopo Piccione
article en

Abstract

Objective. This work aimed at implementing a new recalculation software, ‘RadCalc’ (LAP GmbH Laser Applikationen), independent of the monitor units (MUs) and dosimetric verification of the radiotherapy treatment plans specific for each patient, using the Monte-Carlo (MC) algorithm. Approach. The experimental procedure for the simulation of the treatment plans by the software was started by measuring dosimetric parameters of the operating machine (Elekta Versa HD linear accelerator), subsequently imported into the software. After that, the software tuning on the treatment planning system Monaco v. 5.53 (Elekta) was carried out, which allowed to extract the most suitable Additional Radiation to Light Field Offset parameter to obtain the best performing MC algorithm model for the simulation of the treatment plans. Finally, the above-mentioned model was validated through a statistical analysis comparing the plans recalculated by RadCalc and those experimentally verified on the OCTAVIUS 4D phantom (PTW) and Octavius 1500 (PTW) Detector. Main results. The obtained results showed a good statistical agreement both in the tuning and in the validation part of the MC algorithm model used. The effectiveness of using the RadCalc software was further demonstrated by the high accuracy of the recalculation of the MUs. Significance. It should be noted that the secondary recalculation of the dose, in accordance with Italian legislation on patient radiation protection is one of the actions that the medical physics specialist must ‘prepare for the evaluation and correct verification of the doses administered to patients’ . In this context, it has to be emphasized that a multicenter study group is being set up, within the Italian Association of Medical and Health Physics (AIFM), which brings together the users of RadCalc, aimed at exchanging experiences and improving the performance of the aforementioned software.

Biomedical Physics & Engineering ExpressVol. 12(6)
Azienda ospedaliera "Bianchi-Melacrino-Morelli" (IT)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon