Diagnostic radiology is a cornerstone of modern clinical medicine; however, it constitutes a primary artificial source of ionising radiation exposure to the general population. Maintaining high diagnostic image quality while adhering to radiation protection standards relies fundamentally on periodic quality control (QC) testing of radiographic systems. This study evaluates the operational performance of conventional diagnostic X-ray machines across selected public healthcare facilities in Misurata city by assessing light-to-X-ray field congruence (collimation alignment) and central ray beam perpendicularity. Six diagnostic X-ray units distributed across four healthcare centres were tested using standardized commercial beam and collimator alignment tools at a source-to-image receptor distance (SID) of 100 cm. Compliance was benchmarked against international regulatory tolerances: ≤ 2% of the SID for single-axis field congruence, < 4% for combined orthogonal axes, and < 3° for central ray angular deviation. All six evaluated units achieved a 100% pass rate across both tests. Single-axis displacements ranged from -1.50 cm to +0.50 cm (within the 2.0 cm limit), while combined orthogonal deviations ranged from 1.08 cm to 3.35 cm (within the 4.0 cm limit). Furthermore, central ray angular deviations remained within acceptable thresholds (0.5° to < 3°). These findings confirm robust mechanical calibration and high geometric accuracy across the surveyed units, directly supporting diagnostic reliability and upholding the ALARA (As Low as Reasonably Achievable) principle by minimizing patient radiation doses from unnecessary retakes and improper collimation.

Authors

Publication Details

Journal
AlQalam Journal of Medical and Applied Sciences
Published
2026-11-19
DOI
https://doi.org/10.54361/ajmas.261142
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Quality Control (Field Congruence and Beam Perpendicularity Tests) For Diagnostic X-Ray Machines in Misurata City

Ali Ghabag
AlQalam Journal of Medical and Applied Sciences
Radiation Dose and Imaging
article

Quality Control (Field Congruence and Beam Perpendicularity Tests) For Diagnostic X-Ray Machines in Misurata City

Ali Ghabag
article en

Abstract

Diagnostic radiology is a cornerstone of modern clinical medicine; however, it constitutes a primary artificial source of ionising radiation exposure to the general population. Maintaining high diagnostic image quality while adhering to radiation protection standards relies fundamentally on periodic quality control (QC) testing of radiographic systems. This study evaluates the operational performance of conventional diagnostic X-ray machines across selected public healthcare facilities in Misurata city by assessing light-to-X-ray field congruence (collimation alignment) and central ray beam perpendicularity. Six diagnostic X-ray units distributed across four healthcare centres were tested using standardized commercial beam and collimator alignment tools at a source-to-image receptor distance (SID) of 100 cm. Compliance was benchmarked against international regulatory tolerances: ≤ 2% of the SID for single-axis field congruence, < 4% for combined orthogonal axes, and < 3° for central ray angular deviation. All six evaluated units achieved a 100% pass rate across both tests. Single-axis displacements ranged from -1.50 cm to +0.50 cm (within the 2.0 cm limit), while combined orthogonal deviations ranged from 1.08 cm to 3.35 cm (within the 4.0 cm limit). Furthermore, central ray angular deviations remained within acceptable thresholds (0.5° to < 3°). These findings confirm robust mechanical calibration and high geometric accuracy across the surveyed units, directly supporting diagnostic reliability and upholding the ALARA (As Low as Reasonably Achievable) principle by minimizing patient radiation doses from unnecessary retakes and improper collimation.

AlQalam Journal of Medical and Applied Sciences
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon