RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach

Accurate assessment of radiofrequency electromagnetic field (RF-EMF) induced heating risks for patients with implantable medical electrodes during magnetic resonance imaging (MRI) remains challenging, particularly when implementing the Tier 4(-) evaluation methodology, outlined in the latest ISO 10974 IS1 Committee Draft for Vote (CDV),, in heterogeneous tissue environments. This work introduces an equivalent lumped element model (LEM) to streamline in vivo RF response modeling of implantable medical electrodes in complicated tissue environments without substantially increasing computational burden. Extensive numerical studies were conducted for generic helical implantable medical electrodes with lead lengths of 400, 500, and 600 mm in homogeneous, heterogeneous, and anatomically realistic tissue environments from the Virtual Population model Ella, at 1.5 T and 3.0 T. The LEM-based approach consistently yielded more accurate estimates of RF-EMF-induced power deposited at the implant hotspot than transfer functions derived from simplified tissue configurations representative of Tier 3 evaluations. In particular, the LEM reproduced the equivalent impedance and transfer function of the detailed implantable medical electrode, enabling efficient evaluation of RF-EMF-induced heating in clinically relevant heterogeneous tissue environment. Experimental validation with a commercial implantable medical electrode further showed close agreement between the LEM-predicted transfer function and measurements of the commercial device. These results indicate that the proposed methodology addresses key limitations of conventional modeling strategies and provides a robust, computationally efficient framework for Tier 4(-) RF-EMF-induced heating risk assessment in clinical MRI exposure.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-11-19
DOI
https://doi.org/10.1088/1361-6560/aec4ba
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach

Aiping Yao, Jing Wang, Jinting Fu, Pengfei Yang
Physics in Medicine and Biology
Advanced MRI Techniques and Applications
article

RF-EMF response modeling of implantable medical electrodes under 1.5T and 3T MRI exposure using a lumped element model approach

Aiping Yao, Jing Wang, Jinting Fu, Pengfei Yang
article en

Abstract

Accurate assessment of radiofrequency electromagnetic field (RF-EMF) induced heating risks for patients with implantable medical electrodes during magnetic resonance imaging (MRI) remains challenging, particularly when implementing the Tier 4(-) evaluation methodology, outlined in the latest ISO 10974 IS1 Committee Draft for Vote (CDV),, in heterogeneous tissue environments. This work introduces an equivalent lumped element model (LEM) to streamline in vivo RF response modeling of implantable medical electrodes in complicated tissue environments without substantially increasing computational burden. Extensive numerical studies were conducted for generic helical implantable medical electrodes with lead lengths of 400, 500, and 600 mm in homogeneous, heterogeneous, and anatomically realistic tissue environments from the Virtual Population model Ella, at 1.5 T and 3.0 T. The LEM-based approach consistently yielded more accurate estimates of RF-EMF-induced power deposited at the implant hotspot than transfer functions derived from simplified tissue configurations representative of Tier 3 evaluations. In particular, the LEM reproduced the equivalent impedance and transfer function of the detailed implantable medical electrode, enabling efficient evaluation of RF-EMF-induced heating in clinically relevant heterogeneous tissue environment. Experimental validation with a commercial implantable medical electrode further showed close agreement between the LEM-predicted transfer function and measurements of the commercial device. These results indicate that the proposed methodology addresses key limitations of conventional modeling strategies and provides a robust, computationally efficient framework for Tier 4(-) RF-EMF-induced heating risk assessment in clinical MRI exposure.

Physics in Medicine and Biology
Nanchang University (CN), National Medical Products Administration (CN)
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon