A transcutaneous energy transfer system (TETS) offers a route to eventually eliminating the need for a percutaneous driveline, the principal source of infection in left ventricular assist device (LVAD) therapy, but localised tissue heating still limits clinical translation. ISO 14708-1 restricts the local tissue temperature rise in an active implantable device to ΔT ≤ 2 °C, a margin that high-power pulsed radio frequency (RF) transmission approaches at clinically useful, implanted Li-ion battery charging rates. Establishing whether a given pulsed transmission protocol is thermally safe has so far depended on in vivo measurements. This study investigates a practical method for calibrating a thermostatically controlled, actively circulated isotonic saline (0.9% w/v) in a testbed system environment against the in vivo thermal profiles previously reported in porcine models for a two-channel 200 kHz pulsed RF TETS. The calibration process follows a comprehensive three-phase approach encompassing dry baseline characterisation, wet calibration, and extended 4 h validation at a calibrated saline flow rate of 1.667 L/min, selected from three pump settings. Under the 50 V/320 ms/5 s protocol, the testbed reproduced the porcine sawtooth morphology, a peak secondary coil ΔT of 2.824 ± 0.20 °C and a charging rate of 2494 J/h (±16%, k = 2), comparable to, though not demonstrated equivalent to, the in vivo benchmark at equal transmitter voltage. Its mean plateau ΔT was 0.7–1.2 °C lower, and its post-cessation relaxation four to six times faster than in vivo, indicating that the testbed removes heat more effectively than perfused tissue; fluid flow rate is proposed as the calibration parameter controlling this bias, pending a study with continuously adjustable flow. A 4 h session at 36 °C showed no heat accumulation and kept the array-mean secondary ΔT below 2 °C; however, the thermistor placed next to the secondary coil centre reached 2.27 °C, thus, in this aspect, the protocol did not meet the ISO 14708-1 local limit.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2026-11-19
DOI
https://doi.org/10.3390/s26227326
Primary Topic
Wireless Power Transfer Systems
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Testbed Calibration for Developing Wireless Power Solutions in Heart Failure LVAD Treatment

Omar J. Escalona, DAVID J. MCENEANEY, Mohammad L. Karim, Matthew Hutchinson
Sensors
Wireless Power Transfer Systems
article

Testbed Calibration for Developing Wireless Power Solutions in Heart Failure LVAD Treatment

Omar J. Escalona, DAVID J. MCENEANEY, Mohammad L. Karim, Matthew Hutchinson
article en

Abstract

A transcutaneous energy transfer system (TETS) offers a route to eventually eliminating the need for a percutaneous driveline, the principal source of infection in left ventricular assist device (LVAD) therapy, but localised tissue heating still limits clinical translation. ISO 14708-1 restricts the local tissue temperature rise in an active implantable device to ΔT ≤ 2 °C, a margin that high-power pulsed radio frequency (RF) transmission approaches at clinically useful, implanted Li-ion battery charging rates. Establishing whether a given pulsed transmission protocol is thermally safe has so far depended on in vivo measurements. This study investigates a practical method for calibrating a thermostatically controlled, actively circulated isotonic saline (0.9% w/v) in a testbed system environment against the in vivo thermal profiles previously reported in porcine models for a two-channel 200 kHz pulsed RF TETS. The calibration process follows a comprehensive three-phase approach encompassing dry baseline characterisation, wet calibration, and extended 4 h validation at a calibrated saline flow rate of 1.667 L/min, selected from three pump settings. Under the 50 V/320 ms/5 s protocol, the testbed reproduced the porcine sawtooth morphology, a peak secondary coil ΔT of 2.824 ± 0.20 °C and a charging rate of 2494 J/h (±16%, k = 2), comparable to, though not demonstrated equivalent to, the in vivo benchmark at equal transmitter voltage. Its mean plateau ΔT was 0.7–1.2 °C lower, and its post-cessation relaxation four to six times faster than in vivo, indicating that the testbed removes heat more effectively than perfused tissue; fluid flow rate is proposed as the calibration parameter controlling this bias, pending a study with continuously adjustable flow. A 4 h session at 36 °C showed no heat accumulation and kept the array-mean secondary ΔT below 2 °C; however, the thermistor placed next to the secondary coil centre reached 2.27 °C, thus, in this aspect, the protocol did not meet the ISO 14708-1 local limit.

SensorsVol. 26(22)
University of Ulster (GB), Craigavon Area Hospital (GB)
Wireless Power Transfer Systems
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon