Circumferential Identification of Peripheral Anterior Synechiae by Air-Puff Dynamic Anterior Segment Optical Coherence Tomography

Purpose: The purpose of this study was to objectively and accurately identify peripheral anterior synechiae (PAS) circumferentially in eyes with angle-closure glaucoma (ACG). Methods: Anterior chamber angles (ACAs) were classified as PAS or non-PAS (NPAS) groups by indentation gonioscopy. A custom-built air-puff dynamic anterior segment optical coherence tomography (DAS-OCT) system with an innovative 4-leaf clover-scan pattern acquired 10 frames per ACA along 8 radial directions at 45-degree intervals within 65 ms. Dynamic parameters including the differences in angle opening distance at 750 µm from the scleral spur (AOD750) and trabecular-iris space area at 750 µm anterior to the scleral spur (TISA750) before and after air-puff stimulation, were extracted to establish a binary logistic regression model. Model performance was evaluated in the testing set. Results: Analysis of 68 images from 9 healthy eyes confirmed high measurement repeatability (intraclass correlation coefficient [ICC] = 0.821 for AOD750dif; 0.775 for TISA750dif). A total of 143 anterior segment OCT (AS-OCT) images from 20 ACG eyes were analyzed, with 100 assigned to the training set and 43 to the testing set. Significant differences in AOD750dif and TISA750dif were identified between the NPAS and PAS groups (P < 0.001). The dual-parameter model achieved an area under the curve (AUC) of 0.924 in the training set and 0.913 in the testing set, with an overall accuracy of 83.7%. Conclusions: The air-puff DAS-OCT system enables the noninvasive, dynamic, and circumferential identification of PAS, demonstrating high diagnostic performance. Translational Relevance: This non-contact imaging technology offers a quantitative alternative to indentation gonioscopy, potentially improving ACG screening efficiency and guiding clinical management.

Authors

Institutions

Publication Details

Journal
Translational Vision Science & Technology
Published
2026-11-18
DOI
https://doi.org/10.1167/tvst.15.11.17
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Circumferential Identification of Peripheral Anterior Synechiae by Air-Puff Dynamic Anterior Segment Optical Coherence Tomography

Peijuan Wang, 张绍丹, Cheng Lu, Xikai Wei et al.
Translational Vision Science & Technology
Glaucoma and retinal disorders
article

Circumferential Identification of Peripheral Anterior Synechiae by Air-Puff Dynamic Anterior Segment Optical Coherence Tomography

Peijuan Wang, 张绍丹, Cheng Lu, Xikai Wei, Kaixi Zheng, Jianli Ren, Dexi Zhu
article en

Abstract

Purpose: The purpose of this study was to objectively and accurately identify peripheral anterior synechiae (PAS) circumferentially in eyes with angle-closure glaucoma (ACG). Methods: Anterior chamber angles (ACAs) were classified as PAS or non-PAS (NPAS) groups by indentation gonioscopy. A custom-built air-puff dynamic anterior segment optical coherence tomography (DAS-OCT) system with an innovative 4-leaf clover-scan pattern acquired 10 frames per ACA along 8 radial directions at 45-degree intervals within 65 ms. Dynamic parameters including the differences in angle opening distance at 750 µm from the scleral spur (AOD750) and trabecular-iris space area at 750 µm anterior to the scleral spur (TISA750) before and after air-puff stimulation, were extracted to establish a binary logistic regression model. Model performance was evaluated in the testing set. Results: Analysis of 68 images from 9 healthy eyes confirmed high measurement repeatability (intraclass correlation coefficient [ICC] = 0.821 for AOD750dif; 0.775 for TISA750dif). A total of 143 anterior segment OCT (AS-OCT) images from 20 ACG eyes were analyzed, with 100 assigned to the training set and 43 to the testing set. Significant differences in AOD750dif and TISA750dif were identified between the NPAS and PAS groups (P < 0.001). The dual-parameter model achieved an area under the curve (AUC) of 0.924 in the training set and 0.913 in the testing set, with an overall accuracy of 83.7%. Conclusions: The air-puff DAS-OCT system enables the noninvasive, dynamic, and circumferential identification of PAS, demonstrating high diagnostic performance. Translational Relevance: This non-contact imaging technology offers a quantitative alternative to indentation gonioscopy, potentially improving ACG screening efficiency and guiding clinical management.

Translational Vision Science & TechnologyVol. 15(11)
Wenzhou Medical University (CN)
Good health and well-being
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon