Transvenous lead removal only with manual traction without extraction tools

Objective This study aimed to describe our experience with lead extraction using manual traction.Methods Patients with implanted cardiac leads (n = 130) who underwent lead extraction using manual traction without devices specifically manufactured for extraction were reviewed. Lead extraction difficulty (LED), MB, and ELECTRa Registry Outcome Score (EROS) scores were evaluated to predict extraction success, device requirement, and complications.Results A total of 200 leads were extracted from 130 patients. The mean duration since implantation was 4.05 ± 4.07 years. Leads were successfully removed in 91.5% of patients. Extraction devices were required in 27 (20.8%) patients. Complications occurred in five (3.8%) patients (EROS-1: 3 [2.8%], EROS-2: 0, EROS-3: 2 [18.2%]; p = 0.03). Receiver operating characteristic analysis for major complications: LED score—area under the curve (AUC): 0.793 (0.621–0.965), p = 0.02; MB score—AUC: 0.799 (0.648–0.950), p = 0.02. Receiver operating characteristic analysis for device requirement: LED score—AUC: 0.675 (0.566–0.784), p = 0.005; MB score—AUC: 0.701 (0.600–0.802), p = 0.001.Conclusion Lead extraction using manual traction without extraction tools achieved a 91.5% success rate and 3.8% complication rate. The LED and MB scores significantly predicted complication risk and the need for device use. An EROS score of three was associated with a high risk of complications. Manual traction is more likely to be successful in patients with low scores and may be attempted as an initial strategy (NCT07023900).

Authors

Institutions

Publication Details

Journal
Acta cardiologica. Supplementum
Published
2026-11-18
DOI
https://doi.org/10.1080/00015385.2026.2765184
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Transvenous lead removal only with manual traction without extraction tools

Hasan Arı, Mehmet Melek, Selma Arı, Ayşe Sünbül et al.
1 links
Acta cardiologica. Supplementum
Cardiac pacing and defibrillation studies
article

Transvenous lead removal only with manual traction without extraction tools

Hasan Arı, Mehmet Melek, Selma Arı, Ayşe Sünbül, Gökhan Özmen, Ahmet Tütüncü, Sencer Çamcı, Kubra Doğanay, Tahsin Bozat
article en

Abstract

Objective This study aimed to describe our experience with lead extraction using manual traction.Methods Patients with implanted cardiac leads (n = 130) who underwent lead extraction using manual traction without devices specifically manufactured for extraction were reviewed. Lead extraction difficulty (LED), MB, and ELECTRa Registry Outcome Score (EROS) scores were evaluated to predict extraction success, device requirement, and complications.Results A total of 200 leads were extracted from 130 patients. The mean duration since implantation was 4.05 ± 4.07 years. Leads were successfully removed in 91.5% of patients. Extraction devices were required in 27 (20.8%) patients. Complications occurred in five (3.8%) patients (EROS-1: 3 [2.8%], EROS-2: 0, EROS-3: 2 [18.2%]; p = 0.03). Receiver operating characteristic analysis for major complications: LED score—area under the curve (AUC): 0.793 (0.621–0.965), p = 0.02; MB score—AUC: 0.799 (0.648–0.950), p = 0.02. Receiver operating characteristic analysis for device requirement: LED score—AUC: 0.675 (0.566–0.784), p = 0.005; MB score—AUC: 0.701 (0.600–0.802), p = 0.001.Conclusion Lead extraction using manual traction without extraction tools achieved a 91.5% success rate and 3.8% complication rate. The LED and MB scores significantly predicted complication risk and the need for device use. An EROS score of three was associated with a high risk of complications. Manual traction is more likely to be successful in patients with low scores and may be attempted as an initial strategy (NCT07023900).

Acta cardiologica. Supplementum
Giresun University (TR)
Good health and well-being
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon