Photonic fluence in photobiomodulation: the wavelength adjustment increases rather than removes the disparity in incident photon number
Photobiomodulation lacks a dose descriptor permitting comparison across wavelengths, and the most prominent remedy is a wavelength-adjusted dose termed photonic fluence, expressed in Einstein units, now propagating into supportive-care guidance. We show that this correction is dimensionally inverted. Areal photon number is radiant exposure divided by photon energy, whereas the proposed quantity multiplies the two, so the adjustment widens the disparity in delivered photon number rather than removing it, from 2.4-fold to more than fivefold across 447 to 1064 nm on the source study's own values. We further show that the single empirical anomaly motivating the adjustment arises in a study whose reported dose and reported thermometry cannot both be correct, the reported temperature rise exceeding what the stated delivered energy could produce within the irradiated volume under total absorption. The need the scheme addresses is real. The direction of the correction is not.