Evaluation of Implant Placement Accuracy and CBCT–Surface Comparison Residuals in a Static Implant Surgical-Guide Workflow Using Dental Models
This in vitro study evaluated implant-placement accuracy and cone-beam computed tomography (CBCT)–surface comparison residuals in a static surgical-guide workflow. Seven dental models provided 28 model–region cases for surface comparison; 42 implant positions in five models were evaluated after guided placement. Two observers measured depth and angular deviations three times per position. For each position, the six absolute depth readings and six angular-deviation readings (two observers × three repeats) were averaged separately. Mean absolute axial depth error was 0.369 ± 0.211 mm, and mean angular deviation was 1.864 ± 0.972° (mean ± standard deviation). Forty-one positions (97.6%) met both study-specific placement reference criteria of ≤ 1.0 mm and ≤ 3.5°. Between-observer ICC(A,1) was 0.821 (95% CI, 0.690–0.900) for absolute depth error and 0.794 (95% CI, 0.646–0.884) for angular deviation; Observer 2 showed a repeat-order effect for angular measurements. The mean case-level standard deviation of surface residuals was 0.235 mm. No statistically significant correlations were detected between surface-comparison metrics and placement outcomes across the five matched models (exact two-sided permutation P ≥ 0.083). These measurements characterize the workflow under controlled model conditions. The small matched sample limits conclusions about the association between surface residuals and placement accuracy.
Authors
- Dohyung Lim (ORCID: https://orcid.org/0000-0003-3752-7079)
- Jung‐Tae Lee (ORCID: https://orcid.org/0000-0001-5383-3004)
- Ye-Seul Jung
- Bongju Kim
- Dayeon Park
- Sung-Ho Lee
Institutions
- Sejong University (KR)
- Seoul National University Dental Hospital (KR)
Publication Details
- Journal
- Research Square
- Published
- 2026-11-18
- DOI
- https://doi.org/10.21203/rs.3.rs-11453901/v1
- Primary Topic
- Dental Implant Techniques and Outcomes
- Type
- preprint
Funders
- Ministry of Trade, Industry and Energy
- Korea Institute for Advancement of Technology