Effects of sparse array geometry and element arrangement on low-intensity transcranial focused ultrasound

Sparse arrays are used in low-intensity transcranial focused ultrasound, but the effects of element arrangement, aperture and coverage are not well defined. We simulated nine spherical-cap arrays at 250 kHz, combining 50, 60 and 70 mm aperture radii with 40, 50 and 60% coverage (active-element fraction of the cap area), using 8 mm elements, a 100 mm radius of curvature and a fixed nominal drive of 15 W. A scalar hybrid angular spectrum model with straight-ray phase correction was applied to three skull models and three target placements, with lateral steering up to 30 mm. Matched regular arrays, maximum-density regular arrays and added phase-error conditions were included as controls. Increasing coverage produced target-pressure gains close to the square-root scaling expected from element count, with all nine skull-target pairs remaining within 5% of this expectation. Increasing aperture produced smaller foci but less consistent pressure gains. The effect of quasi-random placement, implemented by thinning a regular candidate lattice, was conditional. At 40% coverage and 30 mm steering, quasi-random arrays reduced peak sidelobe ratios by 1.24-5.34 dB relative to matched regular arrays, whereas regular arrays mostly had similar or lower ratios at offsets up to 20 mm. The sidelobe advantage diminished as coverage increased and the regular pitch approached the candidate-lattice pitch. It also depended on the main-lobe exclusion threshold and, at the midbrain target, did not persist under the larger added phase error. Without steering, quasi-random placement increased integrated skull absorption by 10.0-32.9% at 40% coverage. Steering retention varied more strongly across skull-target pairs than across aperture or coverage, whereas focal fragmentation and off-target maxima occurred mostly at the 50 mm aperture. Within the tested design family, arrangement, aperture and coverage should be considered together when selecting arrays for specific steering and focusing requirements. Biological efficacy and thermal safety were not evaluated.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-11-18
DOI
https://doi.org/10.1371/journal.pone.0355643
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Effects of sparse array geometry and element arrangement on low-intensity transcranial focused ultrasound

Ki Soo Kim, Juyoung Park, Geunho Shim
1 links
PLoS ONE
Ultrasound and Hyperthermia Applications
article

Effects of sparse array geometry and element arrangement on low-intensity transcranial focused ultrasound

Ki Soo Kim, Juyoung Park, Geunho Shim
article en

Abstract

Sparse arrays are used in low-intensity transcranial focused ultrasound, but the effects of element arrangement, aperture and coverage are not well defined. We simulated nine spherical-cap arrays at 250 kHz, combining 50, 60 and 70 mm aperture radii with 40, 50 and 60% coverage (active-element fraction of the cap area), using 8 mm elements, a 100 mm radius of curvature and a fixed nominal drive of 15 W. A scalar hybrid angular spectrum model with straight-ray phase correction was applied to three skull models and three target placements, with lateral steering up to 30 mm. Matched regular arrays, maximum-density regular arrays and added phase-error conditions were included as controls. Increasing coverage produced target-pressure gains close to the square-root scaling expected from element count, with all nine skull-target pairs remaining within 5% of this expectation. Increasing aperture produced smaller foci but less consistent pressure gains. The effect of quasi-random placement, implemented by thinning a regular candidate lattice, was conditional. At 40% coverage and 30 mm steering, quasi-random arrays reduced peak sidelobe ratios by 1.24-5.34 dB relative to matched regular arrays, whereas regular arrays mostly had similar or lower ratios at offsets up to 20 mm. The sidelobe advantage diminished as coverage increased and the regular pitch approached the candidate-lattice pitch. It also depended on the main-lobe exclusion threshold and, at the midbrain target, did not persist under the larger added phase error. Without steering, quasi-random placement increased integrated skull absorption by 10.0-32.9% at 40% coverage. Steering retention varied more strongly across skull-target pairs than across aperture or coverage, whereas focal fragmentation and off-target maxima occurred mostly at the 50 mm aperture. Within the tested design family, arrangement, aperture and coverage should be considered together when selecting arrays for specific steering and focusing requirements. Biological efficacy and thermal safety were not evaluated.

PLoS ONEVol. 21(11)
Gachon University (KR), Kyung Hee University (KR)
Korea Institute for Advancement of Technology, Korea Health Industry Development Institute, National Research Foundation of Korea, Ministry of Science and ICT, South Korea, Institute for Information and Communications Technology Promotion
Good health and well-being
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon