Photonic fluence in photobiomodulation: the wavelength adjustment increases rather than removes the disparity in incident photon number

Photobiomodulation lacks a dose descriptor permitting comparison across wavelengths, and the most prominent remedy is a wavelength-adjusted dose termed photonic fluence, expressed in Einstein units, now propagating into supportive-care guidance. We show that this correction is dimensionally inverted. Areal photon number is radiant exposure divided by photon energy, whereas the proposed quantity multiplies the two, so the adjustment widens the disparity in delivered photon number rather than removing it, from 2.4-fold to more than fivefold across 447 to 1064 nm on the source study's own values. We further show that the single empirical anomaly motivating the adjustment arises in a study whose reported dose and reported thermometry cannot both be correct, the reported temperature rise exceeding what the stated delivered energy could produce within the irradiated volume under total absorption. The need the scheme addresses is real. The direction of the correction is not.

Authors

Publication Details

Published
2026-11-18
DOI
https://doi.org/10.1364/opticaopen.35002276
Primary Topic
Laser Applications in Dentistry and Medicine
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Photonic fluence in photobiomodulation: the wavelength adjustment increases rather than removes the disparity in incident photon number

Michael Richard Hamblin, Mark Cronshaw, Steven Parker
Laser Applications in Dentistry and Medicine
preprint

Photonic fluence in photobiomodulation: the wavelength adjustment increases rather than removes the disparity in incident photon number

Michael Richard Hamblin, Mark Cronshaw, Steven Parker
preprint en

Abstract

Photobiomodulation lacks a dose descriptor permitting comparison across wavelengths, and the most prominent remedy is a wavelength-adjusted dose termed photonic fluence, expressed in Einstein units, now propagating into supportive-care guidance. We show that this correction is dimensionally inverted. Areal photon number is radiant exposure divided by photon energy, whereas the proposed quantity multiplies the two, so the adjustment widens the disparity in delivered photon number rather than removing it, from 2.4-fold to more than fivefold across 447 to 1064 nm on the source study's own values. We further show that the single empirical anomaly motivating the adjustment arises in a study whose reported dose and reported thermometry cannot both be correct, the reported temperature rise exceeding what the stated delivered energy could produce within the irradiated volume under total absorption. The need the scheme addresses is real. The direction of the correction is not.

Laser Applications in Dentistry and Medicine
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon