A Cone-Beam Computed Tomography (CBCT)-Derived Analysis of Inter-fraction Shifts and Correction Burden

Background Stereotactic body radiotherapy (SBRT) requires sub-millimeter geometric precision because of the delivery of ablative radiation doses using highly conformal dose distributions and steep dose gradients. Inter-fraction translational and rotational setup deviations may compromise target coverage and increase radiation exposure to adjacent organs at risk. Daily cone-beam computed tomography (CBCT) with six-degrees-of-freedom (6DoF) couch correction enables comprehensive correction of positional errors; however, real-world data on the magnitude and distribution of these deviations remain limited, particularly from low- and middle-income settings. Objective The study aimed to characterize the magnitude and distribution of inter-fraction translational and rotational setup corrections in patients undergoing CBCT-guided SBRT with 6DoF couch correction, and to quantify the positional correction requirements across treatment fractions. Methods A retrospective observational study was conducted in the Department of Radiation Oncology, Swami Rama Himalayan University, Dehradun, India. Patients who underwent SBRT between January 2022 and December 2024 with daily CBCT guidance and 6DoF couch correction were included. Fraction-wise translational deviations along the lateral (X), longitudinal (Y), and vertical (Z) axes, and rotational deviations in pitch, roll, and yaw were extracted from the treatment delivery system. A three-dimensional rotational vector magnitude was calculated as a composite measure of rotational displacement. Descriptive statistical analysis was performed using mean ± standard deviation (SD) to summarize the absolute magnitudes of fraction-level translational and rotational setup corrections. No inferential statistical testing was performed. Results Fourteen patients contributed 48 evaluable CBCT-guided SBRT treatment fractions. The mean absolute translational corrections were 0.21 ± 0.18 cm laterally, 0.27 ± 0.22 cm longitudinally, and 0.24 ± 0.20 cm vertically, with the largest mean correction observed in the longitudinal direction. Mean absolute rotational corrections were 0.48 ± 0.36° for pitch, 0.44 ± 0.33° for roll, and 0.41 ± 0.31° for yaw. The mean three-dimensional rotational vector magnitude was 0.78 ± 0.42°. Conclusion Daily CBCT-guided 6DoF correction was associated with small inter-fraction translational and rotational setup corrections during SBRT. These findings provide institution-specific baseline data for evaluating immobilization and image-guidance workflows and may inform future patient-level assessment of residual setup uncertainty and planning target volume (PTV) margins.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2026-11-18
DOI
https://doi.org/10.7759/cureus.119871
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Cone-Beam Computed Tomography (CBCT)-Derived Analysis of Inter-fraction Shifts and Correction Burden

Rishabh Dobhal, Vipul Nautiyal, Anmol Mahani, Tanvi Mittal et al.
Cureus
Advanced Radiotherapy Techniques
article

A Cone-Beam Computed Tomography (CBCT)-Derived Analysis of Inter-fraction Shifts and Correction Burden

Rishabh Dobhal, Vipul Nautiyal, Anmol Mahani, Tanvi Mittal, Ravi Kant
article en

Abstract

Background Stereotactic body radiotherapy (SBRT) requires sub-millimeter geometric precision because of the delivery of ablative radiation doses using highly conformal dose distributions and steep dose gradients. Inter-fraction translational and rotational setup deviations may compromise target coverage and increase radiation exposure to adjacent organs at risk. Daily cone-beam computed tomography (CBCT) with six-degrees-of-freedom (6DoF) couch correction enables comprehensive correction of positional errors; however, real-world data on the magnitude and distribution of these deviations remain limited, particularly from low- and middle-income settings. Objective The study aimed to characterize the magnitude and distribution of inter-fraction translational and rotational setup corrections in patients undergoing CBCT-guided SBRT with 6DoF couch correction, and to quantify the positional correction requirements across treatment fractions. Methods A retrospective observational study was conducted in the Department of Radiation Oncology, Swami Rama Himalayan University, Dehradun, India. Patients who underwent SBRT between January 2022 and December 2024 with daily CBCT guidance and 6DoF couch correction were included. Fraction-wise translational deviations along the lateral (X), longitudinal (Y), and vertical (Z) axes, and rotational deviations in pitch, roll, and yaw were extracted from the treatment delivery system. A three-dimensional rotational vector magnitude was calculated as a composite measure of rotational displacement. Descriptive statistical analysis was performed using mean ± standard deviation (SD) to summarize the absolute magnitudes of fraction-level translational and rotational setup corrections. No inferential statistical testing was performed. Results Fourteen patients contributed 48 evaluable CBCT-guided SBRT treatment fractions. The mean absolute translational corrections were 0.21 ± 0.18 cm laterally, 0.27 ± 0.22 cm longitudinally, and 0.24 ± 0.20 cm vertically, with the largest mean correction observed in the longitudinal direction. Mean absolute rotational corrections were 0.48 ± 0.36° for pitch, 0.44 ± 0.33° for roll, and 0.41 ± 0.31° for yaw. The mean three-dimensional rotational vector magnitude was 0.78 ± 0.42°. Conclusion Daily CBCT-guided 6DoF correction was associated with small inter-fraction translational and rotational setup corrections during SBRT. These findings provide institution-specific baseline data for evaluating immobilization and image-guidance workflows and may inform future patient-level assessment of residual setup uncertainty and planning target volume (PTV) margins.

Cureus
Swami Rama Himalayan University (IN)
Good health and well-being
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon