Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices

BACKGROUND: Routine quality assurance (QA) of linear accelerators and electronic portal imaging devices (EPIDs) often requires separate devices and acquisition procedures for geometric, image-quality, radiation-field, multileaf collimator (MLC), and dosimetric checks. A multifunctional phantom could consolidate these assessments within a single QA workflow. PURPOSE: To evaluate the week-to-week reproducibility of measurements obtained with MZ Slab, a multifunctional megavoltage EPID phantom, on three linear accelerators with different EPID configurations. METHODS: Measurements were performed weekly for four weeks on TrueBeam STx, Halcyon, and Harmony Pro. Each session comprised three acquisition sets with phantom removal and repositioning between sets. The phantom provided measurements of geometric accuracy, image quality, material-response isotropy, and radiation-field size, together with qualitative MLC light-field checks. Weekly values were summarized as mean and within-session standard deviation, and changes were assessed relative to week 1. Field sizes were compared with open-field EPID images and IC Profiler measurements. Dose-calibrated portal images acquired on TrueBeam were compared with the week 1 image using mean response change and gamma passing rates. RESULTS: Central-marker offset ranged from 0.24 to 0.78 mm across units, with a maximum change from week 1 of 0.29 mm. Marker-distance deviations changed by no more than 0.03 mm. The maximum changes in mean relative contrast transfer function and low-contrast contrast-to-noise ratio were 0.011 (2.6%) and 1.81 (5.0%), respectively, relative to the week 1 baseline. Uniformity deviation and normalized noise changed by up to 0.50 and 0.17 percentage points, and material-specific normalized response by up to 0.014. Field sizes changed by up to 0.56 mm and agreed with open-field EPID measurements within 0.3 mm. IC Profiler values were 0.2-0.9 mm larger than MZ Slab values. MLC leaf ends remained within the printed ± 1 mm tolerance. At week 4, mean portal-image response had increased by 0.77%, with a gamma passing rate of 99.65% at 1%/1 mm. CONCLUSIONS: MZ Slab supported geometric, image-quality, radiation-field, qualitative MLC, and portal-dosimetry assessments within a single-phantom QA workflow. Measurements on three linacs characterized within-session reproducibility and week-to-week consistency over four weeks. Longer observation and deliberate perturbation testing are required to establish action limits and fault-detection sensitivity.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-11-17
DOI
https://doi.org/10.1002/acm2.70913
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices

Jung-In Kim, Bo‐Wi Cheon, Euntaek Yoon, Inbum Lee et al.
Journal of Applied Clinical Medical Physics
Advanced Radiotherapy Techniques
article

Evaluation of a multifunctional phantom for geometric, image quality, and dosimetric constancy of megavoltage electronic portal imaging devices

Jung-In Kim, Bo‐Wi Cheon, Euntaek Yoon, Inbum Lee, Young Jae Jang, Na Hye Kwon, So‐Yeon Park, Minjae Choi, Yoonsuk Huh, Chang Heon Choi
article en

Abstract

BACKGROUND: Routine quality assurance (QA) of linear accelerators and electronic portal imaging devices (EPIDs) often requires separate devices and acquisition procedures for geometric, image-quality, radiation-field, multileaf collimator (MLC), and dosimetric checks. A multifunctional phantom could consolidate these assessments within a single QA workflow. PURPOSE: To evaluate the week-to-week reproducibility of measurements obtained with MZ Slab, a multifunctional megavoltage EPID phantom, on three linear accelerators with different EPID configurations. METHODS: Measurements were performed weekly for four weeks on TrueBeam STx, Halcyon, and Harmony Pro. Each session comprised three acquisition sets with phantom removal and repositioning between sets. The phantom provided measurements of geometric accuracy, image quality, material-response isotropy, and radiation-field size, together with qualitative MLC light-field checks. Weekly values were summarized as mean and within-session standard deviation, and changes were assessed relative to week 1. Field sizes were compared with open-field EPID images and IC Profiler measurements. Dose-calibrated portal images acquired on TrueBeam were compared with the week 1 image using mean response change and gamma passing rates. RESULTS: Central-marker offset ranged from 0.24 to 0.78 mm across units, with a maximum change from week 1 of 0.29 mm. Marker-distance deviations changed by no more than 0.03 mm. The maximum changes in mean relative contrast transfer function and low-contrast contrast-to-noise ratio were 0.011 (2.6%) and 1.81 (5.0%), respectively, relative to the week 1 baseline. Uniformity deviation and normalized noise changed by up to 0.50 and 0.17 percentage points, and material-specific normalized response by up to 0.014. Field sizes changed by up to 0.56 mm and agreed with open-field EPID measurements within 0.3 mm. IC Profiler values were 0.2-0.9 mm larger than MZ Slab values. MLC leaf ends remained within the printed ± 1 mm tolerance. At week 4, mean portal-image response had increased by 0.77%, with a gamma passing rate of 99.65% at 1%/1 mm. CONCLUSIONS: MZ Slab supported geometric, image-quality, radiation-field, qualitative MLC, and portal-dosimetry assessments within a single-phantom QA workflow. Measurements on three linacs characterized within-session reproducibility and week-to-week consistency over four weeks. Longer observation and deliberate perturbation testing are required to establish action limits and fault-detection sensitivity.

Journal of Applied Clinical Medical PhysicsVol. 27(12)
Seoul National University (KR), Seoul National University Hospital (KR), Veterans Health Service Medical Center (KR), Seoul National University Hospital Biomedical Research Institute (KR)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon