Interferometric diameter topography calibration of test spheres
Test spheres as used in the calibration of coordinate measuring machine stylus and other free form measuring devices are typically characterised by a small set of diameters and the sphericity along these diameters. This sparse sampling limits the number of positions where the true diameter is known, reducing the sphere’s utility in calibration. To enhance data density and decrease the uncertainty of calibrations, an extension to PTB’s double-ended interferometer (DEI) (Rau and Schödel 2014 Double-Ended Interferometer for Measuring Gauge Blocks Without Wringing pp 859–62) has been developed (Schödel and Fischedick 2021 Meas. Sci. Technol. 32 084010). This work presents the implementation, measurements and evaluation methods for this extension. First, the DEI including the extension is introduced and the optical setup explained. Then the evaluation methods are presented. The evaluation is performed in two steps. First, pixelwise diameter values are calculated for all measurements performed on a sphere, then, a stitching algorithm is employed to combine a series of measurements to one representation of the sphere surface figure. The evaluation is tested with a measurement series on a 30 mm silicon sphere. The uncertainty budget for a diameter is presented, with a combined standard uncertainty of 20 nm.
Authors
- René Schödel (ORCID: https://orcid.org/0000-0002-7597-9036)
- Guido Bartl (ORCID: https://orcid.org/0000-0002-0746-1200)
- Tillman Neupert-Wentz (ORCID: https://orcid.org/0009-0002-3226-9509)
Institutions
- Physikalisch-Technische Bundesanstalt (DE)
Publication Details
- Journal
- Measurement Science and Technology
- Published
- 2026-11-18
- DOI
- https://doi.org/10.1088/1361-6501/aebef2
- Primary Topic
- Advanced Measurement and Metrology Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00