Effect of radiotherapy on cardiac implantable electronic devices – evaluation of potential explanatory variables in more than a decade

BACKGROUND AND PURPOSE: As the population ages, an increasing number of patients are undergoing radiotherapy with an implanted cardiac device. Therefore, ensuring the safety of these devices during radiotherapy is crucial. This study aims to analyse the effect of radiotherapy on such devices, with a particular focus on changes in lead variables. METHODS: This retrospective study examined all patients with a cardiac implantable electronic device (CIED) who underwent radiotherapy at a single institution between 2012 and 2022. Device function and associated changes in lead parameters were analysed before and after radiotherapy. RESULTS: The study cohort included 279 patients who underwent 310 courses of radiotherapy. During the 11-year study period, one device failure requiring replacement due to a software error was recorded. This occurred following irradiation with 15 MV photons. Changes in individual lead parameters were more common, occurring in almost half of all radiotherapy courses. Lead impedances were particularly sensitive to changes, with high-voltage impedances being affected more frequently (61% of impedances) than other impedances. Sensing, pacing thresholds, and battery voltage were less frequently affected, with changes occurring in fewer than 10% of the analysed leads or batteries. None of the observed changes manifested as clinical symptoms. There was no statistically significant association between radiation parameters and the observed changes in lead values with only a few courses using neutron-generating photon energies. Furthermore, no dose-effect relationship for device doses could be established. Receiving multiple courses of radiotherapy in one year or over several years did not increase the risk to the device or cause greater variation in lead variables. CONCLUSION: Only one safety event occurred during the observation period, resulting in an error rate of 0.4% across 11 years and 279 patients. Changes in individual lead parameters were more frequent. However, no clear association could be established between the observed changes and radiation parameters. In particular, changes in lead variables were not explained by the dose to the device or repeated courses of radiotherapy. Therefore, the risk to the safety and function of cardiac devices from radiotherapy appears to be small.

Authors

Institutions

Publication Details

Journal
Radiation Oncology
Published
2026-11-18
DOI
https://doi.org/10.1186/s13014-026-02940-8
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Effect of radiotherapy on cardiac implantable electronic devices – evaluation of potential explanatory variables in more than a decade

Henrike A. K. Hillmann, Roland Merten, Alexander Christiansen, David Duncker et al.
5 links
Radiation Oncology
Cardiac pacing and defibrillation studies
article

Effect of radiotherapy on cardiac implantable electronic devices – evaluation of potential explanatory variables in more than a decade

Henrike A. K. Hillmann, Roland Merten, Alexander Christiansen, David Duncker, Hans Christiansen, Johanna Müller-Leisse
article en

Abstract

BACKGROUND AND PURPOSE: As the population ages, an increasing number of patients are undergoing radiotherapy with an implanted cardiac device. Therefore, ensuring the safety of these devices during radiotherapy is crucial. This study aims to analyse the effect of radiotherapy on such devices, with a particular focus on changes in lead variables. METHODS: This retrospective study examined all patients with a cardiac implantable electronic device (CIED) who underwent radiotherapy at a single institution between 2012 and 2022. Device function and associated changes in lead parameters were analysed before and after radiotherapy. RESULTS: The study cohort included 279 patients who underwent 310 courses of radiotherapy. During the 11-year study period, one device failure requiring replacement due to a software error was recorded. This occurred following irradiation with 15 MV photons. Changes in individual lead parameters were more common, occurring in almost half of all radiotherapy courses. Lead impedances were particularly sensitive to changes, with high-voltage impedances being affected more frequently (61% of impedances) than other impedances. Sensing, pacing thresholds, and battery voltage were less frequently affected, with changes occurring in fewer than 10% of the analysed leads or batteries. None of the observed changes manifested as clinical symptoms. There was no statistically significant association between radiation parameters and the observed changes in lead values with only a few courses using neutron-generating photon energies. Furthermore, no dose-effect relationship for device doses could be established. Receiving multiple courses of radiotherapy in one year or over several years did not increase the risk to the device or cause greater variation in lead variables. CONCLUSION: Only one safety event occurred during the observation period, resulting in an error rate of 0.4% across 11 years and 279 patients. Changes in individual lead parameters were more frequent. However, no clear association could be established between the observed changes and radiation parameters. In particular, changes in lead variables were not explained by the dose to the device or repeated courses of radiotherapy. Therefore, the risk to the safety and function of cardiac devices from radiotherapy appears to be small.

Radiation OncologyVol. 21(1)
Medizinische Hochschule Hannover (DE)
Boston Scientific Corporation
Good health and well-being
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon