Finite Element-Based Design of Experiments for the Optimal Geometry of a Ti6Al4V Hip Implant Stem: Influence of Loading Direction

Background/Objectives: Finite element optimization of hip implant stems is usually performed under a single simplified load, often directed along the neck axis. This study asks whether the geometric parameters governing stem stress change with the idealized static loading direction. Methods: A Taguchi L27 array of a parametric Ti6Al4V stem (four geometric parameters, three levels each) was analyzed with quadratic tetrahedral elements under 2300 N applied along the neck axis (Case A) and vertically through the femoral head (Case B). The response was the peak von Mises stress outside the potted region and a 12 mm exclusion region at the tied head–trunnion junction. Contributions were quantified by single-factor analysis of variance (ANOVA). Results: The two loads gave different outcomes. Under neck-axis loading, the peak stress was 74.2–92.8 MPa and neck length explained 100.0% of the variance. Under vertical loading, the peak stress was 218.8–636.8 MPa, neck base radius (58.5%) and neck top radius (29.9%) dominated, and neck length explained 4.7%. The effect of neck length reversed: increasing it from 20 to 30 mm lowered the neck-axis stress by 19.8% but raised the mean vertical stress by 15.4%. The rank correlation between cases was weak (ρ = −0.23). Conclusions: Within this model and these factor ranges, optimizing under neck-axis loading alone fixes only the neck length and leaves the vertical response undetermined. Designs 12, 24, and 9 gave the best compromise between the two loads. Stem geometry should therefore be optimized under both loading directions.

Authors

Institutions

Publication Details

Journal
Prosthesis
Published
2026-11-18
DOI
https://doi.org/10.3390/prosthesis8110123
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Finite Element-Based Design of Experiments for the Optimal Geometry of a Ti6Al4V Hip Implant Stem: Influence of Loading Direction

Raviraj Shetty, Adithya Lokesh Hegde, J P Supriya, Gururaj Bolar et al.
1 links
Prosthesis
Orthopaedic implants and arthroplasty
article

Finite Element-Based Design of Experiments for the Optimal Geometry of a Ti6Al4V Hip Implant Stem: Influence of Loading Direction

Raviraj Shetty, Adithya Lokesh Hegde, J P Supriya, Gururaj Bolar, Vinyas, Harishkumar Kamat, Amulya Shetty, Jonathan Monteiro
article en

Abstract

Background/Objectives: Finite element optimization of hip implant stems is usually performed under a single simplified load, often directed along the neck axis. This study asks whether the geometric parameters governing stem stress change with the idealized static loading direction. Methods: A Taguchi L27 array of a parametric Ti6Al4V stem (four geometric parameters, three levels each) was analyzed with quadratic tetrahedral elements under 2300 N applied along the neck axis (Case A) and vertically through the femoral head (Case B). The response was the peak von Mises stress outside the potted region and a 12 mm exclusion region at the tied head–trunnion junction. Contributions were quantified by single-factor analysis of variance (ANOVA). Results: The two loads gave different outcomes. Under neck-axis loading, the peak stress was 74.2–92.8 MPa and neck length explained 100.0% of the variance. Under vertical loading, the peak stress was 218.8–636.8 MPa, neck base radius (58.5%) and neck top radius (29.9%) dominated, and neck length explained 4.7%. The effect of neck length reversed: increasing it from 20 to 30 mm lowered the neck-axis stress by 19.8% but raised the mean vertical stress by 15.4%. The rank correlation between cases was weak (ρ = −0.23). Conclusions: Within this model and these factor ranges, optimizing under neck-axis loading alone fixes only the neck length and leaves the vertical response undetermined. Designs 12, 24, and 9 gave the best compromise between the two loads. Stem geometry should therefore be optimized under both loading directions.

ProsthesisVol. 8(11)
Manipal Academy of Higher Education (IN), Father Muller Medical College Hospital (IN), Visvesvaraya Technological University (IN)
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon