A Time-Correlated Single-Photon Counting Instrument: Timing Resolution, Count Rate and Linearity Budgets from Computed Timing Physics

Measuring how long a molecule glows after a flash of light needs an instrument that times each emitted photon against the flash, and the timing must be precise enough to resolve lifetimes of a few nanoseconds. Operating principle: a time-correlated single-photon counting instrument, in which a pulsed laser excites a sample and a fast detector with a time-to-digital converter records the delay of each emitted photon, building a decay histogram one photon at a time. The counting instrument reaches a timing jitter of 40 picoseconds at a wavelength of 780 nanometres, with a time bin width of 10 picoseconds and a maximum count rate of 10 million counts per second. The instrument times single photons to forty picoseconds and builds decay curves from their distribution, which is the precision that makes fluorescence lifetime imaging and time-resolved spectroscopy possible. SIMULATED, NOT MEASURED. This is a computed study for a civilian sensing purpose: the device serves imaging, sensing, communications or illumination, and it carries no targeting or weapon application against any person, state or terrestrial object.

Authors

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2026-11-18
DOI
https://doi.org/10.5281/zenodo.24336849
Primary Topic
Advanced Fluorescence Microscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Time-Correlated Single-Photon Counting Instrument: Timing Resolution, Count Rate and Linearity Budgets from Computed Timing Physics

William Alexander Aldred
4 links
Zenodo (CERN European Organization for Nuclear Research)
Advanced Fluorescence Microscopy Techniques
article

A Time-Correlated Single-Photon Counting Instrument: Timing Resolution, Count Rate and Linearity Budgets from Computed Timing Physics

William Alexander Aldred
article en

Abstract

Measuring how long a molecule glows after a flash of light needs an instrument that times each emitted photon against the flash, and the timing must be precise enough to resolve lifetimes of a few nanoseconds. Operating principle: a time-correlated single-photon counting instrument, in which a pulsed laser excites a sample and a fast detector with a time-to-digital converter records the delay of each emitted photon, building a decay histogram one photon at a time. The counting instrument reaches a timing jitter of 40 picoseconds at a wavelength of 780 nanometres, with a time bin width of 10 picoseconds and a maximum count rate of 10 million counts per second. The instrument times single photons to forty picoseconds and builds decay curves from their distribution, which is the precision that makes fluorescence lifetime imaging and time-resolved spectroscopy possible. SIMULATED, NOT MEASURED. This is a computed study for a civilian sensing purpose: the device serves imaging, sensing, communications or illumination, and it carries no targeting or weapon application against any person, state or terrestrial object.

Zenodo (CERN European Organization for Nuclear Research)
Advanced Fluorescence Microscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon