Monte Carlo assessment of a D–T neutron generator-based beam shaping assembly for BNCT: moderator effects on neutron spectrum and dose

This study presents a Monte Carlo parametric assessment of a multilayer beam shaping assembly (BSA) for boron neutron capture therapy using a 14.1 MeV D–T neutron source. A validated reference calculation was retained for the principal beam-quality quantities, while additional controlled calculations were used to reassess material substitution, finite-source and aperture effects, and one-at-a-time layer-thickness sensitivity. Under a common computational framework, AlF 3 provided the highest epithermal response and the lowest fast-to-epithermal ratio among the tested moderator substitutions. Paraffin and light water reduced the epithermal component by approximately 74% and 73%, respectively, relative to AlF 3 , whereas D 2 O and polystyrene produced intermediate and similar responses. The validated reference configuration yielded an epithermal neutron flux of 1.1×10 9 n cm⁻ 2 s⁻ 1 and a beam directionality of 0.78; however, its fast-neutron dose per epithermal flux, 2.4×10⁻ 13 Gy·cm 2 /n, remained about 20% above the historical IAEA-TECDOC-1223 design benchmark. Additional calculations showed that the finite 1 cm-radius source did not produce a statistically resolved change relative to the point-source reference, whereas increasing the aperture from 10 to 24 cm increased the aperture-averaged epithermal response. Thickness variations revealed substantial transmission trade-offs and did not establish a unique global optimum. The 36 cm AlF 3 , 4 cm Fe, and 2.6 cm Cd dimensions are therefore reported as the reference geometry of the existing assembly. The results support continued computational refinement but do not constitute clinical qualification.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-11-18
DOI
https://doi.org/10.1038/s41598-026-80838-x
Primary Topic
Boron Compounds in Chemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Monte Carlo assessment of a D–T neutron generator-based beam shaping assembly for BNCT: moderator effects on neutron spectrum and dose

Aydin Ghalehasadi, S. Asghari, A. Alavi, M. Ghapanvary et al.
Scientific Reports
Boron Compounds in Chemistry
article

Monte Carlo assessment of a D–T neutron generator-based beam shaping assembly for BNCT: moderator effects on neutron spectrum and dose

Aydin Ghalehasadi, S. Asghari, A. Alavi, M. Ghapanvary, V. Dehghani
article en

Abstract

This study presents a Monte Carlo parametric assessment of a multilayer beam shaping assembly (BSA) for boron neutron capture therapy using a 14.1 MeV D–T neutron source. A validated reference calculation was retained for the principal beam-quality quantities, while additional controlled calculations were used to reassess material substitution, finite-source and aperture effects, and one-at-a-time layer-thickness sensitivity. Under a common computational framework, AlF 3 provided the highest epithermal response and the lowest fast-to-epithermal ratio among the tested moderator substitutions. Paraffin and light water reduced the epithermal component by approximately 74% and 73%, respectively, relative to AlF 3 , whereas D 2 O and polystyrene produced intermediate and similar responses. The validated reference configuration yielded an epithermal neutron flux of 1.1×10 9 n cm⁻ 2 s⁻ 1 and a beam directionality of 0.78; however, its fast-neutron dose per epithermal flux, 2.4×10⁻ 13 Gy·cm 2 /n, remained about 20% above the historical IAEA-TECDOC-1223 design benchmark. Additional calculations showed that the finite 1 cm-radius source did not produce a statistically resolved change relative to the point-source reference, whereas increasing the aperture from 10 to 24 cm increased the aperture-averaged epithermal response. Thickness variations revealed substantial transmission trade-offs and did not establish a unique global optimum. The 36 cm AlF 3 , 4 cm Fe, and 2.6 cm Cd dimensions are therefore reported as the reference geometry of the existing assembly. The results support continued computational refinement but do not constitute clinical qualification.

Scientific Reports
University of Sistan and Baluchestan (IR), University of Tabriz (IR), Nuclear Science and Technology Research Institute (IR), Fasa University (IR)
Good health and well-being
Boron Compounds in Chemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon