Comparison of refractive error measurements obtained with a pyramid wavefront sensor topo-aberrometer and an auto kerato-refractometer against subjective refraction

Clinical relevance Refraction is a crucial part of an ocular examination. Ensuring the accuracy and efficiency of objective baseline measurements is essential for optimising clinical workflow and improving final prescription outcomes.Background Agreement between objective refraction of Topcon KR-800 auto kerato-refractometer and CSO Osiris-T topo-aberrometer with subjective refraction (SR), has never been evaluated before. This study aimed to evaluate the agreement and clinical interchangeability of these devices with SR.Methods This observational cross-sectional study included patients with best corrected distance visual acuity ≥6/7.5. The proportions of spherical equivalent (SE) differences within ±0.25 D and ±0.50 D were reported descriptively. SE, cylinder, axis, J0 and J45 were compared, using Bland-Altman plots with 95% limits of agreement, and double-angle astigmatic plots of vectorial cylinder difference.Results The sample included 118 eyes of 60 women and 58 men with a mean (SD) age of 35.6 years (11.4), range 18 to 64 years. The percentages of differences between KR-800 and Osiris-T with SR were, respectively, 68.97% and 29.31% within ±0.25 D, and 92.24% and 59.48% within ±0.50 D. The mean spherical equivalent difference with SR was −0.09 D for KR-800 and 0.33 D for Osiris-T, and the 95% limits of agreement were −0.68 to 0.51 D and −0.54 to 1.21 D, respectively. The centroid of the vectorial astigmatic differences with SR was 0.04 D at 72º for KR-800 and 0.24 D at 41º for Osiris-T. Mean axis difference with SR was 2.47º (95% limits of agreement = -10.28 to 15.23º) for KR-800, and −2.73º (95% limits of agreement = -33.22 to 27.76º) for Osiris-T.Conclusion Both devices provide a useful reference, but they are not interchangeable with SR. KR-800 demonstrated closer agreement with SR across all parameters.

Authors

Institutions

Publication Details

Journal
Clinical and Experimental Optometry
Published
2026-11-18
DOI
https://doi.org/10.1080/08164622.2026.2765422
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Comparison of refractive error measurements obtained with a pyramid wavefront sensor topo-aberrometer and an auto kerato-refractometer against subjective refraction

Gemma Julio, Àlex Julien, Rafael Ignacio Barraquer, Fathima Junaidha
Clinical and Experimental Optometry
Ophthalmology and Visual Impairment Studies
article

Comparison of refractive error measurements obtained with a pyramid wavefront sensor topo-aberrometer and an auto kerato-refractometer against subjective refraction

Gemma Julio, Àlex Julien, Rafael Ignacio Barraquer, Fathima Junaidha
article en

Abstract

Clinical relevance Refraction is a crucial part of an ocular examination. Ensuring the accuracy and efficiency of objective baseline measurements is essential for optimising clinical workflow and improving final prescription outcomes.Background Agreement between objective refraction of Topcon KR-800 auto kerato-refractometer and CSO Osiris-T topo-aberrometer with subjective refraction (SR), has never been evaluated before. This study aimed to evaluate the agreement and clinical interchangeability of these devices with SR.Methods This observational cross-sectional study included patients with best corrected distance visual acuity ≥6/7.5. The proportions of spherical equivalent (SE) differences within ±0.25 D and ±0.50 D were reported descriptively. SE, cylinder, axis, J0 and J45 were compared, using Bland-Altman plots with 95% limits of agreement, and double-angle astigmatic plots of vectorial cylinder difference.Results The sample included 118 eyes of 60 women and 58 men with a mean (SD) age of 35.6 years (11.4), range 18 to 64 years. The percentages of differences between KR-800 and Osiris-T with SR were, respectively, 68.97% and 29.31% within ±0.25 D, and 92.24% and 59.48% within ±0.50 D. The mean spherical equivalent difference with SR was −0.09 D for KR-800 and 0.33 D for Osiris-T, and the 95% limits of agreement were −0.68 to 0.51 D and −0.54 to 1.21 D, respectively. The centroid of the vectorial astigmatic differences with SR was 0.04 D at 72º for KR-800 and 0.24 D at 41º for Osiris-T. Mean axis difference with SR was 2.47º (95% limits of agreement = -10.28 to 15.23º) for KR-800, and −2.73º (95% limits of agreement = -33.22 to 27.76º) for Osiris-T.Conclusion Both devices provide a useful reference, but they are not interchangeable with SR. KR-800 demonstrated closer agreement with SR across all parameters.

Clinical and Experimental Optometry
Barraquer Opthalmology Center (ES)
Universitat Autònoma de Barcelona
Good health and well-being
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon