OpenBench CM1: an open-source double-beam LED colorimeter, and a calibration-free dilution test showing that spectral bandwidth, not noise, limits its linear range
Low-cost LED colorimeters are usually validated with a calibration line over a narrow range, so the limits of their linearity are rarely measured. I present OpenBench CM1, a 3D-printed double-beam colorimeter built for about £13, in which two BPW34 photodiodes, transimpedance amplifiers and a 16-bit ADC measure the sample beam and the LED simultaneously, with LED on/off subtraction in every reading. A built-in 34-minute self-characterisation measured a single-reading precision of 0.89 mAU (limit of detection 2.9 mAU) and showed that the reference ratio removes source drift (sample channel −0.035 ± 0.008% min⁻¹; ratio −0.001 ± 0.005% min⁻¹, p = 0.88). A 13-step two-fold dilution series of Brilliant Blue FCF (E133) spanning an 8,192-fold concentration range then showed that accuracy, not precision, limits the instrument. Absorbance obeyed Beer–Lambert (log–log slope 0.92 ± 0.04, residuals ≤ 2 mAU) only below about 0.09 AU. Above this, every two-fold dilution changed the reading by a constant factor of 1.155 ± 0.007 (log–log slope 0.21), so a visually opaque stock read 0.835 AU rather than the ≈155 AU expected. An opaque-cuvette test limited stray light to below 0.004%, excluding the usual stray-light explanation. The behaviour is instead a polychromatic deviation: the detected band (white LED, red gel filter and a silicon photodiode that remains sensitive into the near-infrared) is wider than the dye's absorption band. Because the two-fold ratio needs no standards of known concentration, it gives any builder a simple test of an instrument's linear range, and the results specify a narrow-band 630 nm source for the next version. Design files, firmware and data: https://github.com/taharafi2010/openbench-cm1 (archived at https://doi.org/10.5281/zenodo.24323813). Project site: https://openbench.oesimracing.com
Authors
- Taha Rafi
Publication Details
- Journal
- Zenodo (CERN European Organization for Nuclear Research)
- Published
- 2026-11-17
- DOI
- https://doi.org/10.5281/zenodo.24323896
- Primary Topic
- Color Science and Applications
- Type
- preprint