The Pressure Behavior of a Digital Drainage System

Objective: Digital chest drainage systems are widely used since they provide objective monitoring of air leak and may shorten drainage duration and hospital stay. However, the pressure behavior of these systems is unknown. This study aimed to evaluate whether the preset suction pressure of the Thopaz+ digital drainage system (Medela AG, Baar, Switzerland) corresponds to the actual measured intrapleural pressure during initial activation and also compares it with an analog water-seal system. Methods: An in vitro model simulating the pleural space was developed using a ventilator, a test lung, and a sealed barrel containing a balloon to mimic lung expansion. Two configurations were tested: spontaneous breathing and mechanical ventilation. Intrapleural pressure was measured while the Thopaz+ system was set to −8 cm H 2 O suction. Additional experiments simulated a pneumothorax with adjustable air leaks (150 and 550 mL/min). Also, a 2-step protocol (−5 cm H 2 O followed by −8 cm H 2 O) was evaluated. Results were also compared with an analog water-seal drainage system. Results: In both spontaneous and ventilated models, the Thopaz+ system showed a transient pressure overshoot during the first minutes. With a preset suction of −8 cm H 2 O, the measured intrapleural pressure reached −27 cm H 2 O. This negative overshoot decreased with increasing air leak. A gradual 2-step suction approach prevented pressures from dropping below −8 cm H 2 O. In contrast, the analog drainage system did not generate pressures below the preset level. Conclusions: The Thopaz+ digital drainage system generates transient overshoot in negative intrapleural pressures immediately after activation. The clinical relevance of this phenomenon requires further investigation.

Authors

Institutions

Publication Details

Journal
Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
Published
2026-11-17
DOI
https://doi.org/10.1177/15569845261498020
Primary Topic
Pleural and Pulmonary Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

The Pressure Behavior of a Digital Drainage System

Ron G. H. Speekenbrink, Alberdina W. Klomp, Michiel Wagenaar, Frans H.C. de Jongh
1 links
Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
Pleural and Pulmonary Diseases
article

The Pressure Behavior of a Digital Drainage System

Ron G. H. Speekenbrink, Alberdina W. Klomp, Michiel Wagenaar, Frans H.C. de Jongh
article en

Abstract

Objective: Digital chest drainage systems are widely used since they provide objective monitoring of air leak and may shorten drainage duration and hospital stay. However, the pressure behavior of these systems is unknown. This study aimed to evaluate whether the preset suction pressure of the Thopaz+ digital drainage system (Medela AG, Baar, Switzerland) corresponds to the actual measured intrapleural pressure during initial activation and also compares it with an analog water-seal system. Methods: An in vitro model simulating the pleural space was developed using a ventilator, a test lung, and a sealed barrel containing a balloon to mimic lung expansion. Two configurations were tested: spontaneous breathing and mechanical ventilation. Intrapleural pressure was measured while the Thopaz+ system was set to −8 cm H 2 O suction. Additional experiments simulated a pneumothorax with adjustable air leaks (150 and 550 mL/min). Also, a 2-step protocol (−5 cm H 2 O followed by −8 cm H 2 O) was evaluated. Results were also compared with an analog water-seal drainage system. Results: In both spontaneous and ventilated models, the Thopaz+ system showed a transient pressure overshoot during the first minutes. With a preset suction of −8 cm H 2 O, the measured intrapleural pressure reached −27 cm H 2 O. This negative overshoot decreased with increasing air leak. A gradual 2-step suction approach prevented pressures from dropping below −8 cm H 2 O. In contrast, the analog drainage system did not generate pressures below the preset level. Conclusions: The Thopaz+ digital drainage system generates transient overshoot in negative intrapleural pressures immediately after activation. The clinical relevance of this phenomenon requires further investigation.

Innovations Technology and Techniques in Cardiothoracic and Vascular Surgery
Medisch Spectrum Twente (NL), Max Planck University of Twente Center for Complex Fluid Dynamics (NL), University of Twente (NL)
Good health and well-being
Pleural and Pulmonary Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon