Autonomous Robotic Respiratory Motion Compensation and Needle Insertion via MRI Correspondence Models for CT‐Guided Hepatic Percutaneous Procedures
BACKGROUND: Respiratory-induced organ motion limits the precision of hepatic percutaneous procedures, while needle misplacement may cause misdiagnosis, incomplete treatment, or damage to healthy tissue. METHODS: We present an autonomous motion-model-based approach for respiratory motion compensation and needle insertion. The models estimate lesion position from an external surrogate signal correlated with respiratory motion. They were trained using MRI data and evaluated in a hybrid operating room using optical tracking as the surrogate. The framework was compared with a baseline method using electromagnetic tracking. Validation involved 15 insertions into a liver phantom at different breath-hold positions. RESULTS: The overall insertion error was 3.3 mm, differing by less than 1 mm from the baseline and demonstrating acceptable accuracy for hepatic percutaneous procedures. CONCLUSIONS: The proposed approach enables accurate needle placement during respiratory motion and may reduce the risks of misdiagnosis, incomplete treatment, and healthy-tissue damage.
Authors
- Momen Abayazid (ORCID: https://orcid.org/0000-0003-1857-5921)
- Ana Cordón Avila (ORCID: https://orcid.org/0009-0003-2148-5457)
Institutions
- King Fahd University of Petroleum and Minerals (SA)
- University of Twente (NL)
Publication Details
- Journal
- International Journal of Medical Robotics and Computer Assisted Surgery
- Published
- 2026-11-16
- DOI
- https://doi.org/10.1002/rcs.70252
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00