An experimental imaging dose estimation method for the phase-II proton CT scanner

PURPOSE: Proton computed tomography (pCT) improves range accuracy in proton therapy by directly mapping proton stopping power relative to water (RSP). Imaging dose remains a crucial factor in determining the feasibility of daily image guidance. This study presents an experimental methodology for estimating imaging dose in the phase-II pCT scanner, utilizing phase space data recorded at the front tracker to eliminate phantom-specific corrections. The estimated doses are validated against ionization chamber (IC) measurements. METHODS: Dose measurements were conducted at multiple locations and angles within two phantoms, with data acquisition triggered by either the front trackers or the energy detector. Baseline scans were also performed without phantoms. Monte Carlo (MC) simulations were used to estimate imaging dose in single projections and full scans. Eight front-tracker-triggered baseline datasets were processed for proton track reconstruction and fluence estimation. After assessing inter-run variability, phase space files were averaged to generate an input source for MC simulations. The simulated imaging doses were renormalized to the averaged datasets and quantitatively compared against IC measurements. RESULTS: Analyzed proton track numbers exhibited consistency across datasets, with coefficients of variation below 2.0%. Estimated doses agreed with IC measurements within ±1.8% for the CTDI phantom; head phantom underestimated -4.6% (single projection) and -6.7% (full scan). Uncertainties from multiple measurements remained below 1.9%. CONCLUSION: This study introduces a reliable workflow for imaging dose estimation in the phase-II pCT scanner. The approach demonstrates satisfactory agreement with IC measurements, supporting its feasibility for evaluating imaging dose in image-guided and adaptive proton therapy.

Authors

Institutions

Publication Details

Journal
Physica Medica
Published
2026-11-17
DOI
https://doi.org/10.1016/j.ejmp.2026.107223
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

An experimental imaging dose estimation method for the phase-II proton CT scanner

Katia Parodi, Reinhard W. Schulte, Mark Pankuch, Guillaume Landry et al.
1 links
Physica Medica
Radiation Therapy and Dosimetry
article

An experimental imaging dose estimation method for the phase-II proton CT scanner

Katia Parodi, Reinhard W. Schulte, Mark Pankuch, Guillaume Landry, Robert P. Johnson, Jonathan Frederik Bortfeldt, George Dedes, Lei Chen
article en

Abstract

PURPOSE: Proton computed tomography (pCT) improves range accuracy in proton therapy by directly mapping proton stopping power relative to water (RSP). Imaging dose remains a crucial factor in determining the feasibility of daily image guidance. This study presents an experimental methodology for estimating imaging dose in the phase-II pCT scanner, utilizing phase space data recorded at the front tracker to eliminate phantom-specific corrections. The estimated doses are validated against ionization chamber (IC) measurements. METHODS: Dose measurements were conducted at multiple locations and angles within two phantoms, with data acquisition triggered by either the front trackers or the energy detector. Baseline scans were also performed without phantoms. Monte Carlo (MC) simulations were used to estimate imaging dose in single projections and full scans. Eight front-tracker-triggered baseline datasets were processed for proton track reconstruction and fluence estimation. After assessing inter-run variability, phase space files were averaged to generate an input source for MC simulations. The simulated imaging doses were renormalized to the averaged datasets and quantitatively compared against IC measurements. RESULTS: Analyzed proton track numbers exhibited consistency across datasets, with coefficients of variation below 2.0%. Estimated doses agreed with IC measurements within ±1.8% for the CTDI phantom; head phantom underestimated -4.6% (single projection) and -6.7% (full scan). Uncertainties from multiple measurements remained below 1.9%. CONCLUSION: This study introduces a reliable workflow for imaging dose estimation in the phase-II pCT scanner. The approach demonstrates satisfactory agreement with IC measurements, supporting its feasibility for evaluating imaging dose in image-guided and adaptive proton therapy.

Physica MedicaVol. 152
University of California, Santa Cruz (US), Loma Linda University (US), LMU Klinikum (DE), Northwestern Medicine (US), Ludwig-Maximilians-Universität München (DE)
China Scholarship Council
Good health and well-being
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon