Mechatronic Design, Biomechanical Analysis, and Kinematic Validation of a Continuous Passive Motion Slider–Crank–Lead Screw System for Lower-Limb Rehabilitation: From Post-Stroke Therapy to Potential Post-Spaceflight Reconditioning

This study developed and preliminarily validated a robotic Continuous Passive Motion (CPM) system intended for lower-limb rehabilitation in post-stroke adult patients with hemiplegia. Building upon a previous redesign and simulation stage using a virtual patient model, an adjustable structure made of AISI 316L stainless steel was fabricated, integrating the mechanical, electronic, and control subsystems for passive knee and ankle mobilization through a lead screw mechanism driven by a NEMA23 stepper motor, a TB6600 driver, and an ESP32 microcontroller operating under an open-loop feedforward control scheme. As a preliminary stage prior to clinical evaluation, functional validation was conducted through three knee therapy tests, two with a healthy adult volunteer and one under unloaded conditions, using different angular ranges (90–130°, 30–26°, and 50–70°). The device successfully executed the programmed movements without mechanical or electronic failures, completing all three repetitions in each session and recording observed differences ranging from 2.00% to 11.54% between the programmed and measured joint angles. These results demonstrate the preliminary technical feasibility of the proposed prototype. The present validation was limited to a single adult volunteer and three knee-motion tests. Future work will include validation with volunteers representing different anthropometric profiles (1.55–1.75 m), verification of emergency stop response times below 100 ms, and the incorporation of encoder-based position feedback as preliminary steps toward subsequent clinical evaluation.

Authors

Institutions

Publication Details

Journal
Technologies
Published
2026-11-16
DOI
https://doi.org/10.3390/technologies14110757
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Mechatronic Design, Biomechanical Analysis, and Kinematic Validation of a Continuous Passive Motion Slider–Crank–Lead Screw System for Lower-Limb Rehabilitation: From Post-Stroke Therapy to Potential Post-Spaceflight Reconditioning

Ricardo Palomares, José Cornejo, Mariela Vargas, Jeanette Borja Arroyo et al.
Technologies
Prosthetics and Rehabilitation Robotics
article

Mechatronic Design, Biomechanical Analysis, and Kinematic Validation of a Continuous Passive Motion Slider–Crank–Lead Screw System for Lower-Limb Rehabilitation: From Post-Stroke Therapy to Potential Post-Spaceflight Reconditioning

Ricardo Palomares, José Cornejo, Mariela Vargas, Jeanette Borja Arroyo, Adrian Nacarino, Carol Sandoval, Josue Alata-Rey, Cesar Martel, Margarita Murillo, Harnnol Suclupe
article en

Abstract

This study developed and preliminarily validated a robotic Continuous Passive Motion (CPM) system intended for lower-limb rehabilitation in post-stroke adult patients with hemiplegia. Building upon a previous redesign and simulation stage using a virtual patient model, an adjustable structure made of AISI 316L stainless steel was fabricated, integrating the mechanical, electronic, and control subsystems for passive knee and ankle mobilization through a lead screw mechanism driven by a NEMA23 stepper motor, a TB6600 driver, and an ESP32 microcontroller operating under an open-loop feedforward control scheme. As a preliminary stage prior to clinical evaluation, functional validation was conducted through three knee therapy tests, two with a healthy adult volunteer and one under unloaded conditions, using different angular ranges (90–130°, 30–26°, and 50–70°). The device successfully executed the programmed movements without mechanical or electronic failures, completing all three repetitions in each session and recording observed differences ranging from 2.00% to 11.54% between the programmed and measured joint angles. These results demonstrate the preliminary technical feasibility of the proposed prototype. The present validation was limited to a single adult volunteer and three knee-motion tests. Future work will include validation with volunteers representing different anthropometric profiles (1.55–1.75 m), verification of emergency stop response times below 100 ms, and the incorporation of encoder-based position feedback as preliminary steps toward subsequent clinical evaluation.

TechnologiesVol. 14(11)
Florida Institute of Technology (US), Universidad Ricardo Palma (PE), St Martins Hospital (GB), Universidad Continental (PE), Universidad Politécnica de Madrid (ES), Capitol Technology University (US)
Prosthetics and Rehabilitation Robotics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon