Apparent membrane lung dysfunction during veno-arterial extracorporeal membrane oxygenation following methylene blue administration: Case report
infective endocarditis with embolic hepatic abscesses. Severe postoperative vasoplegia was treated with repeated methylene blue boluses and infusions while on central veno-arterial extracorporeal membrane oxygenation (V-A ECMO) for refractory cardiogenic and vasoplegic shock. Following multiple methylene blue administrations, the ECMO return tubing appeared dark, raising concern for membrane lung dysfunction (MLD). However, post-membrane lung blood gases confirmed adequate oxygen exchange and visual inspection on the post-membrane lung side of the circuit revealed bright red blood within the membrane lung (ML) and all rigid connectors. The discoloration was determined to have resulted from methylene blue staining of the flexible PVC tubing which mimicked the appearance of the post-ML flexible PVC tubing in a failed membrane lung despite the oxygenator functioning appropriately. Awareness of this artifact is critical to avoid unnecessary interventions; evaluation of post-ML blood gases and visual comparison with unstained ECMO circuit components can distinguish this benign finding from true MLD.
Authors
- Benjamin Hohlfelder (ORCID: https://orcid.org/0000-0001-8452-3519)
- Chase M. Donaldson (ORCID: https://orcid.org/0000-0003-0274-6159)
- Nichol Hirz
- Marie Pacetti
- Sheena Settlemires
- Tyler Newbauer
- Taylor Barry
Institutions
- Cleveland Clinic (US)
Publication Details
- Journal
- Perfusion
- Published
- 2026-11-17
- DOI
- https://doi.org/10.1177/02676591261500978
- Primary Topic
- Mechanical Circulatory Support Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.00