Comparative biomechanical study of a novel variable diameter circular external fixation device ( VDCF ) for stabilizing a simulated tibial fracture gap in bovine cadaver specimens

OBJECTIVE: To compare the biomechanical performance of a novel variable diameter circular external fixation device (VDCF) with a conventional circular external fixator (CEF) using a standardized osteotomized buffalo tibia model under axial compression and torsional loading. STUDY DESIGN: Experimental in vitro biomechanical comparative study. SAMPLE POPULATION: A total of 36 adult bovine tibiae obtained postmortem from buffaloes. METHODS: The tibiae were randomly allocated into three groups (n = 12 each): VDCF, CEF, and intact bone. A standardized mid-diaphyseal transverse osteotomy with a 20 mm gap was created to simulate an unstable fracture model. Constructs were subjected to axial compression and torsional testing to failure using displacement-controlled protocols. Key biomechanical parameters evaluated included failure load, axial stiffness, and maximum torque derived from load-deformation curves. Statistical analysis was performed using one-way ANOVA followed by Tukey's HSD test. RESULTS: Intact tibiae demonstrated significantly greater failure load and axial stiffness than both fixation constructs (p < .05). No significant differences were detected between VDCF and CEF in failure load or axial stiffness (p > .05). During torsional testing, maximum torque did not differ significantly between constructs (p > .05); however, VDCF exhibited significantly greater torsional stiffness than CEF (p < .05). Failure modes were primarily characterized by plastic deformation and wire bending in both constructs, while fixation-bolt loosening was observed only in the CEF group. CONCLUSION: The VDCF showed no significant differences compared with conventional circular external fixation in failure load, axial stiffness, maximum torque, and torsional strength, while exhibiting significantly greater torsional stiffness under the experimental conditions. CLINICAL SIGNIFICANCE/IMPACT: The VDCF represents a potential alternative fixation configuration with practical advantages in construct adaptability and cost for fracture stabilization in large animals.

Authors

Institutions

Publication Details

Journal
Veterinary Surgery
Published
2026-11-16
DOI
https://doi.org/10.1111/vsu.70177
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Comparative biomechanical study of a novel variable diameter circular external fixation device ( VDCF ) for stabilizing a simulated tibial fracture gap in bovine cadaver specimens

Kuskur Sannappa Naik Kamalesh Kumar, ABHIJIT MOTIRAM PAWDE, Obli Rajendran Vinodhkumar, Rekha Pathak et al.
Veterinary Surgery
Bone fractures and treatments
article

Comparative biomechanical study of a novel variable diameter circular external fixation device ( VDCF ) for stabilizing a simulated tibial fracture gap in bovine cadaver specimens

Kuskur Sannappa Naik Kamalesh Kumar, ABHIJIT MOTIRAM PAWDE, Obli Rajendran Vinodhkumar, Rekha Pathak, Abhishek Chandra Saxena, T. Sai Kumar, Aman Kumar Tiwari, Amarpal, Rohit Kumar
article en

Abstract

OBJECTIVE: To compare the biomechanical performance of a novel variable diameter circular external fixation device (VDCF) with a conventional circular external fixator (CEF) using a standardized osteotomized buffalo tibia model under axial compression and torsional loading. STUDY DESIGN: Experimental in vitro biomechanical comparative study. SAMPLE POPULATION: A total of 36 adult bovine tibiae obtained postmortem from buffaloes. METHODS: The tibiae were randomly allocated into three groups (n = 12 each): VDCF, CEF, and intact bone. A standardized mid-diaphyseal transverse osteotomy with a 20 mm gap was created to simulate an unstable fracture model. Constructs were subjected to axial compression and torsional testing to failure using displacement-controlled protocols. Key biomechanical parameters evaluated included failure load, axial stiffness, and maximum torque derived from load-deformation curves. Statistical analysis was performed using one-way ANOVA followed by Tukey's HSD test. RESULTS: Intact tibiae demonstrated significantly greater failure load and axial stiffness than both fixation constructs (p < .05). No significant differences were detected between VDCF and CEF in failure load or axial stiffness (p > .05). During torsional testing, maximum torque did not differ significantly between constructs (p > .05); however, VDCF exhibited significantly greater torsional stiffness than CEF (p < .05). Failure modes were primarily characterized by plastic deformation and wire bending in both constructs, while fixation-bolt loosening was observed only in the CEF group. CONCLUSION: The VDCF showed no significant differences compared with conventional circular external fixation in failure load, axial stiffness, maximum torque, and torsional strength, while exhibiting significantly greater torsional stiffness under the experimental conditions. CLINICAL SIGNIFICANCE/IMPACT: The VDCF represents a potential alternative fixation configuration with practical advantages in construct adaptability and cost for fracture stabilization in large animals.

Veterinary Surgery
Indian Veterinary Research Institute (IN)
Bone fractures and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon