Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation

BACKGROUND: Defibrillation efficacy of the subcutaneous implantable cardioverter-defibrillator (S-ICD) depends on shock vector geometry and tissue impedances influenced by implantation technique. Whether longitudinal low-voltage impedance (LVI) measurements differ across contemporary implantation strategies, and whether these differences are explained by quantitative radiographic geometry, remains incompletely characterized. OBJECTIVE: To determine whether S-ICD implantation technique is associated with longitudinal LVI trajectories and whether continuous radiographic implant geometry provides information beyond categorical PRAETORIAN risk classification. METHODS: We performed a single-center retrospective cohort study of 136 adults receiving an S-ICD between 2014 and 2024, classified by operative documentation as subcutaneous (SC; n = 27), fluoroscopy-marked intermuscular (FIM; n = 73), or fluoroscopy-marked ultrasound-guided intermuscular (FUSIM; n = 36) techniques. Per-device LVI was summarized overall and across prespecified windows (0-30 days, 6 months-1 year, 2-5 years). Technique differences were analyzed. We fit a prespecified mixed-effects interaction model that included coil-sternum distance, scaled in 5 mm increments, and assessed correlations between LVI and radiographic measures. RESULTS: In paired analyses, LVI increased by a mean of 13.3 Ω between the 0-30-day and 2-5-year windows. Overall LVI differed by implantation technique (Kruskal-Wallis p = 0.019), driven by higher values in SC versus FIM (Holm-adjusted p = 0.041) and FUSIM (p = 0.022). Additional impedance differences across techniques were absent in the 0-30-day window but emerged during follow-up. In the prespecified interaction mixed-effects model, SC exhibited a significant additional rise relative to FUSIM during the 2-5-year window (interaction β = 7.82 Ω, 95% CI 0.98-14.67; p = 0.025). Greater coil-sternum distance was independently associated with higher LVI (13.6 Ω per cm, 95% CI 4.2-23.0; p = 0.005) and correlated with chronic LVI (n = 76; r = 0.441; Benjamini-Hochberg q < 0.001). The PRAETORIAN score was computable in 76 devices and classified 84% as low risk; within that low-risk stratum, coil-sternum distance remained associated with chronic LVI (ρ = 0.320, p = 0.010) whereas the score did not (ρ = 0.136, p = 0.282), and each 5 mm of coil-sternum distance corresponded to 10.7 Ω higher LVI after adjustment for BMI (95% CI 2.68-18.73, p = 0.010). High-voltage shock impedance at implant correlated strongly with chronic LVI (n = 99; r = 0.71, p < 0.001), and the association persisted over 2-5 years (n = 69; r = 0.61, p < 0.001). CONCLUSION: Longitudinal LVI varies with S-ICD implantation strategy, with time-dependent divergence between SC and intermuscular implants. Within the low-risk PRAETORIAN category, where the categorical score has no dynamic range, continuously measured coil-sternum distance remains associated with chronic impedance. Continuous implant geometry and longitudinal impedance complement, rather than replace, radiographic scoring of implant position.

Authors

Institutions

Publication Details

Journal
Journal of Cardiovascular Electrophysiology
Published
2026-11-17
DOI
https://doi.org/10.1111/jce.70568
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation

Albert Joseph Rogers, Nitish Badhwar, Fabián Sanchis‐Gomar, Sanjiv M. Narayan et al.
3 links
Journal of Cardiovascular Electrophysiology
Cardiac pacing and defibrillation studies
article

Implantation Technique, Continuous Implant Geometry, and Longitudinal Low‐Voltage Impedance After Subcutaneous Implantable Cardioverter‐Defibrillator Implantation

Albert Joseph Rogers, Nitish Badhwar, Fabián Sanchis‐Gomar, Sanjiv M. Narayan, Tina Baykaner, Leah Schloss, Paul J. Wang, Marco Valentin Perez, Mohan N. Viswanathan, Alexander C. Perino, Roy M. John
article en

Abstract

BACKGROUND: Defibrillation efficacy of the subcutaneous implantable cardioverter-defibrillator (S-ICD) depends on shock vector geometry and tissue impedances influenced by implantation technique. Whether longitudinal low-voltage impedance (LVI) measurements differ across contemporary implantation strategies, and whether these differences are explained by quantitative radiographic geometry, remains incompletely characterized. OBJECTIVE: To determine whether S-ICD implantation technique is associated with longitudinal LVI trajectories and whether continuous radiographic implant geometry provides information beyond categorical PRAETORIAN risk classification. METHODS: We performed a single-center retrospective cohort study of 136 adults receiving an S-ICD between 2014 and 2024, classified by operative documentation as subcutaneous (SC; n = 27), fluoroscopy-marked intermuscular (FIM; n = 73), or fluoroscopy-marked ultrasound-guided intermuscular (FUSIM; n = 36) techniques. Per-device LVI was summarized overall and across prespecified windows (0-30 days, 6 months-1 year, 2-5 years). Technique differences were analyzed. We fit a prespecified mixed-effects interaction model that included coil-sternum distance, scaled in 5 mm increments, and assessed correlations between LVI and radiographic measures. RESULTS: In paired analyses, LVI increased by a mean of 13.3 Ω between the 0-30-day and 2-5-year windows. Overall LVI differed by implantation technique (Kruskal-Wallis p = 0.019), driven by higher values in SC versus FIM (Holm-adjusted p = 0.041) and FUSIM (p = 0.022). Additional impedance differences across techniques were absent in the 0-30-day window but emerged during follow-up. In the prespecified interaction mixed-effects model, SC exhibited a significant additional rise relative to FUSIM during the 2-5-year window (interaction β = 7.82 Ω, 95% CI 0.98-14.67; p = 0.025). Greater coil-sternum distance was independently associated with higher LVI (13.6 Ω per cm, 95% CI 4.2-23.0; p = 0.005) and correlated with chronic LVI (n = 76; r = 0.441; Benjamini-Hochberg q < 0.001). The PRAETORIAN score was computable in 76 devices and classified 84% as low risk; within that low-risk stratum, coil-sternum distance remained associated with chronic LVI (ρ = 0.320, p = 0.010) whereas the score did not (ρ = 0.136, p = 0.282), and each 5 mm of coil-sternum distance corresponded to 10.7 Ω higher LVI after adjustment for BMI (95% CI 2.68-18.73, p = 0.010). High-voltage shock impedance at implant correlated strongly with chronic LVI (n = 99; r = 0.71, p < 0.001), and the association persisted over 2-5 years (n = 69; r = 0.61, p < 0.001). CONCLUSION: Longitudinal LVI varies with S-ICD implantation strategy, with time-dependent divergence between SC and intermuscular implants. Within the low-risk PRAETORIAN category, where the categorical score has no dynamic range, continuously measured coil-sternum distance remains associated with chronic impedance. Continuous implant geometry and longitudinal impedance complement, rather than replace, radiographic scoring of implant position.

Journal of Cardiovascular Electrophysiology
Cardiovascular Institute of the South (US), Stanford Medicine (US), Stanford Cardiovascular Institute (US), Universidad de Burgos (ES), Stanford University (US)
Good health and well-being
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon